TM 11-6625-1576-15

Section V
Paragraphs 5-30 to 5-32

Model 333A/334A

1)

2)

3)

5-30.

a.

NOTE

If the 60 on the Frequency dial is not under
the cursor after step j is completed, refer
to Figure 7-1 and perform the following

steps:

Remove set screws (#2) from knob on Frequency
dial, and then remove knob (#31). Do not change
C4 setting.

Loosen retaining screws on Frequency dial (#3),
and slip dial until 60 is under cursor, being
careful not to disturb C4 setting.

Tighten retaining screws and replace knob and
set screws.

VOLTMETER GAIN ADJUSTMENTS.

Connect Voltmeter Calibrator (-hp- Model

738BR) to Distortion Analyzer as shown in Figure 5-5.

Remove shorting bar between earth ground and floating
ground terminals.

b. Set Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . VOLTMETER

METER RANGE Selector . . . .0.001 VOLTS

c. Set Voltmeter Calibrator to 400 ˜ RMS, 1.0

millivolt output.

d. Adjust A2R34 (400 cps CAL. 2) for a meter in-

dication of 1.0.

e. Set Voltmeter Calibrator to 400 ~ RMS, O. 3

millivolt output. Set Distortion Analyzer METER
RANGE selector to 0.0003 VOLTS.

f. The meter shall indicate 3.0 ± 1/3 of a minor

division (±O. 55%). If necessary, change A2R30 to set

reading within these limits,

NOTE

If transistors are replaced in the meter

amplifier, it may be necessary to change
A2R30. Typical values for this resistor
are between 140-260 ohms. Increasing
resistance decreases the meter reading
and vice versa.

g. Perform the Voltmeter Accuracy Check of Para-

graph 5-17. The meter shall indicate within the
limits specified.

5-31. VOLTMETER FREQUENCY RESPONSE

ADJUSTMENT.

a. Connect the test equipment to the 333A/334A as

shown in Figure 5-6.

NOTE

Range switch cover must be in place be-
fore performing these adjustments.

b. Set the Distortion Analyzer controls as follows:

FUNCTION Selector . . . . . . . VOLTMETER

METER RANGE Selector . . . 0.0003 VOLTS

c. Set the frequency response test set for 400 cps

ad adjust the output amplitude for a Distortion Ana-
lyzer meter indication of 0.9 on the 1.0 scale.

5-10

d. Adjust the frequency response test set meter

control for a set level indication

e. Change frequency response test set frequency t'

20 cps.

f. Readjust the amplitude until the frequency re-

sponse test set meter indicates set level.

g. The Distortion Analyzer meter shall indicate

0.9 ± 5%.

NOTE

If reading exceeds these limits, change
A2C25. Typical value for this capacitor
is between 50 and 200 µf. Increasing ca-
pacitance reduces the meter reading, and

vice versa.

h. Switch Distortion Analyzer METER RANGE

selector to O. 001 VOLTS, Repeat steps c and d.

j. Change frequency response test set to 5 cps.

Repeat step f.

k. The Distortion Analyzer meter shall indicate

0.9 ± 5%.

m. Switch the Distortion Analyzer METER RANGE

selector to O. 01 VOLTS. Repeat steps c and d.

n. Adjust frequency response test set frequency

for 2Mc and adjust the output amplitude until the fre-

quency response test set indicates set level.

p. Adjust A2R39 (GAIN ADJUST) for a Distortion

Analyzer meter indication of O. 9 on the 1.0 scale.

q. Change the frequency response test set frequen-

cy to 3 Mc. Readjust the output amplitude until the
frequency response test set indicates set level.

r. Adjust A2C21 (3 Mc CAL trimmer) for a Dis-

tortion Analyzer meter Indication of O. 9 on the 1.0
scale.

NOTE

The frequency response adjustments made
thus far interact, It will be necessary to

repeat steps n thru r until all adjustments

are within specifications.

s. Switch Distortion Analyzer METER RANGE to

O. 3 VOLTS. Repeat steps c and d.

t. Change the frequency response test set frequen-

cy to 2.5 Mc. Readjust the output amplitude until the
meter indicates set level. Adjust S3C6 (VOLTAGE
CAL at 2.5 Mc O. 3 v RANGE trimmer) for a Distor-
tion Analyzer meter indication of O. 9 on the 1.0 scale.

v. Switch the Distortion Analyzer METER RANGE

selector to 3 VOLTS. Repeat steps c and d.

w. Change the frequency response test set frequen-

cy to 200 Kc. Repeat step f.

x. Adjust S3C1 (200 Kc CAL) for a Distortion Ana-

lyzer indication of O. 9 on the 1.0 scale.

5-32. SENSITIVITY SWITCH CALIBRATION.

a. Connect instruments as shown in Figure 5-6.

Summary of Contents for 333A

Page 1: ...1 6625 1576 15 DEPARTMENT OF THE ARMY TECHNICAL MANUAL ORGANIZATIONAL DS GS AND DEPOT MAINTENANCE MANUAL DISTORTION ANALYZER HEWLETT PACKARD MODELS 333A AND 334A HEADQUARTERS DEPARTMENT OF THE ARMY MAY 1967 ...

Page 2: ...TM 11 6625 1576 15 WARNING DANGEROUS VOLTAGES EXIST IN THIS EQUIPMENT Be careful when working on the power supply and on the 115 volt ac line connections DO NOT TAKE CHANCES ...

Page 3: ...stortion Measurement in AM Carriers 4 1 4 8 Voltmeter Operation 4 1 4 10 Schematic Theory 4 1 4 11 Impedance Converter Circuit 4 1 4 14 Rejection Amplifier Circuit 4 2 4 36 High Pass Filter 4 5 4 38 Meter Circuit 4 5 4 46 Power Supply Circuit 4 7 4 51 RF Detector Circuit 334A only l 4 7 Section Page V MAINTENANCE 5 1 5 l Introduction 5 1 5 2 Test Equipment Required 5 1 5 5 Performance Checks 5 1 S...

Page 4: ......

Page 5: ...TM 11 6625 1576 15 Model 333A 334A Section 1 Figure 1 1 and Table 1 1 Figure 1 1 Model 333A Distortion Analyzer 1 1 Table 1 1 Specifications ...

Page 6: ...TM 11 6625 1576 15 Section I Model 333A 334A Table 1 1 Table 1 1 Specifications Cont d 1 2 ...

Page 7: ...t h r o u g h p u b l i c a t i o n s s u p p l y c h a n n e l s T h e i n d e x l i s t s t h e i n d i v i d u a l p a r t s 1 0 2 0 3 5 P e t c a n d t h e l a t e s t c h a n g e s t o a n d r e v i s i o n s o f e a c h e q u i p m e n t p u b l i c a t i o n l A 3 F o r m s a n d R e c o r d s a R e p o r t s o f M a i n t e n a n c e a n d U n s a t i s f a c t o r y E q u i p m e n t U s ...

Page 8: ......

Page 9: ...s a fre quency range of 5 cps to 3 Mc 20 cps to 500 Kc for 300 pv range and a voltage range of 300 pv to 300 v rms full scale 1 5 The AM detector included in the Model 334A is a broadband dc restoring peak detector consisting of a semiconductor diode and filter circuit AM distortion levels as low as O 3 can be measured on a 3 v to 8 v rms carrier modulated 30 in the standard broadcast band and low...

Page 10: ......

Page 11: ...pin on the power cable three prong connec tor is the ground wire 2 9 To preserve the protection feature when operat ing the instrument from a two contact outlet use a three prong to two prong adapter and connect the green pigtail on the adapter to ground 2 10 INSTALLATION 2 11 The Model 333A 334A is fully transistorized therefore no special cooling is required However the instrument should not be ...

Page 12: ...TM 11 6625 1576 15 Section III Model 333A 334A Figure 3 1 Figure 3 1 Front and Rear Panel Description 3 0 ...

Page 13: ...ver age to effective values in a true sine wave is approxi mately O 9 to 1 When the meter is used to measure complex waves the voltage indicated may not be the rms value of the signal applied This deviation of meter indication exists because the ratios of average to effective values are usually not the same in a com plexwave as in a sine wave The amount of deviation depends on magnitude and phase ...

Page 14: ... SET LEVEL position thus making the next range 30 etc h Adjust SENSITIVITY VENIER control for full scale deflection j Position FRIQUENCY RANGE selector and frequency dial to fundamental frequency of input signal 3 2 Model 333A 334A k Position FUNCTION selector to DISTORTION m Adjust METER RANGE selector and frequency dial vernier control for minimum meter indication n Adjust COARSE and FINE BALANC...

Page 15: ...tion III Paragraphs 3 23 to 3 27 3 25 The 333A and 334A perform as general purpose AC Voltmeters when the FUNCTION selector is set to VOLTMETER position NOTE With the FUNCTION selector in VOLT METER position the SENSITIVITY se lector is disabled a Position METER RANGE selector to a range exceeding the value of the signal to be measured b Connect signal to INPUT terminals c Select a METER RANGE to ...

Page 16: ...F i g u r e 4 1 B l o c k D i a g r a m T M 1 1 6 6 2 5 1 5 7 6 1 5 S e c t i o n I V F i g u r e 4 1 Model 333A 334A ...

Page 17: ...ost attenuator is used to limit the input signal level applied to the metering circuit to 1 mv for full scale deflectlon The metering circuit sensitivity is increased to 300 uV for full scale deflection on the 300 pv range The metering circuit provides a visual indication of the OPERATION distortion level of the input signal In addition to the visual indication provided by the meter the OUTPUT ter...

Page 18: ...g the bridge output using the input signal as a reference 4 20 When the Wien bridge is not tuned exactly to the frequency to be nulled a portion of the fundamental frequency will appear at the bridge output The phase of this signal depends on which leg of the bridge is not tuned or on the relative errors in tuning if neither is set correctly The magnitude of the signal is propor tional to the magn...

Page 19: ...of the FREQUENCY RANGE switch S4 switches A5R56 in parallel with A5R55 on the top three frequency ranges A6DS2 will become brighter and lower the resistance of A6V2 through A6V5 making variation in resistance less than on the two lower ranges However less variation in resistance is needed to tune the leg because the impedance in the reactive leg becomes progressively less as the higher frequency r...

Page 20: ...tector A5Q4 4 30 Refer to Figure 4 5 simplified partial schematic for detector operation The discussion is applicable to both resistive and reactive detector circuits 4 31 The signals from the error amplifier A5Q2 and A5Q3 will be equal and of opposite phase and will cancel out each other when the detector A5Q4 is off However when the positive half of the reference square wave gates A5Q4 on the si...

Page 21: ...tivity is increased to 300 uv for full scale deflection on the 300 uV range by switching resistors A2R29 and A2R30 into the calibration network Resis tor A2R41 and capacitor A2C29 are also switched into the calibration network on the 300 Pv range to extend the passband of the amplifier 4 42 METER AMPLIFIER CIRCUIT 4 43 The meter amplifier circuit consists of a five stage amplifier circuit A2Q5 thr...

Page 22: ...TM 11 6625 1576 15 Section IV Figure 4 7 Model 333A 334A Figure 4 7 Bandwidth Versus Null Depth 4 6 ...

Page 23: ...nal series regulator type and operates the same as the 25 volt regulated supply 4 50 Diodes A1CR5 and A1CR6 are coupling and pro tection diodes for external battery supplies The diodes protect the series regulator circuits from application of incorrect polarity at the battery input terminals The diodes also protect external batteries from being charged when the ac power is being used with batterie...

Page 24: ...TM 11 6625 1576 15 Section V Table 5 1 Model 333A 334A Table 5 1 Test Equipment Required 5 0 ...

Page 25: ...as follows MIN Position 1 next step Position 2 etc to Max Position 6 5 9 FUNDAMENTAL REJECTION CHECK a b S e t Connect 33lA 332A as shown in Figure 5 1 Set Distortion Analyzer controls as follows FUNCTION Selector VOLTMETER METER RANGE Selector 1 VOLT FREQUENCY RANGE Selector X1OO Frequency Dial 50 filter White Instr Lab Model 2640 5Kc on Distortion Analyzer meter d Switch Distortion Analyzer FUNC...

Page 26: ...control for a O db indication on the Distortion Analyzer meter h Switch Distortion Analyzer FUNCTION selector to DISTORTION The meter reading shall not change more than 0 6 db Model 333A 334A j Repeat steps a through h at the frequency settings indicated in Table 5 2 The meter readings shall change within the limits specified Table 5 2 Second Harmonic Accuracy Check 5 11 a b c DISTORTION INTRODUCE...

Page 27: ...t Check 5 12 FREQUENCY CALIBRATION ACCURACY CHECK b c TM 11 6625 1576 15 Section V Paragraph 5 12 Table 5 3 and Figure 5 3 NOTE From 5 cps to 10 cps the FREQUENCY dial may be as much as 3 low In this test the dial is held constant and the in put frequency is varied and monitored If the dial is low the input frequency at null will be high If the period of the in put frequency is measured it will be...

Page 28: ...eter shall indicate less than 30 pf d Switch the Distortion Analyzer on the 0 3 range and measure capacitance Meter shall indicate less than 60 pf 5 4 e Set Distortion Analyzer controls as follows FUNCTION Selector DISTORTION SENSITIVITY Selector MIN METER RANGE Selector VOLTS f Measure Capacitance at each SENSITIVITY selector setting of the Distortion Analyzer The L C meter shall indicate less th...

Page 29: ...ut c Set Distortion Analyzer FUNCTION Selector to VOLTMETER d Set voltmeter calibrator for 400 cps output e Check the Distortion Analyzer voltmeter full scale readings on all ranges against the appropriate rms input voltages from the voltmeter calibrator The voltmeter accuracy shall be within 2 f Set the Distortion Analyzer METER RANGE selector to 1 VOLT range Transformer set to 105 v 125 v 5 18 H...

Page 30: ...l indicate 0 9 the toler ances indicated k Switch the Frequency Response Test Set RANGE SELECTOR to the 1 3 Mc position Set the FREQ TUNING dial to the frequencies listed in Table 5 5 Adjust the Frequency Response Test Set AMPLITUDE control until the meter indicates SET LEVEL after each frequency setting The Distortlon Analyzer meter shall indicate 0 9 the tolerances indicated 5 6 n Repeat steps b...

Page 31: ... 33 Adjustment locations are shown in Figure 5 8 5 24 The Adjustment and Calibration Procedure is performed with the ac power cord connected to nomi nal line voltage 115 v 230 v 50 to 1000 cps MODE switch to MANUAL HIGH PASS FILTER to OUT and NORM RF DET switch to NORM unless otherwise specified 5 25 METER MECHANICAL ZERO SET 5 26 The meter is properly zero set when the pointer rests over the zero...

Page 32: ...TM11 6625 1576 15 Section V Model 333A 334A Figure 5 8 Figure 5 8 Component and Adjustment Location 5 8 ...

Page 33: ... Figure 5 9 g Repeat steps b through f adjusting A3R30 in stead of A3R16 5 29 BRIDGE BALANCE ADJUSTMENT C3 MAINTAIN THE LEAD DRESS TO C3 AND TO THE TUNING CAPACITOR C4 ANY CHANGE IN LEAD DRESS WILL CAUSE A CHANGE IN CAPACITANCE a Connect the Test Oscillator to the Distortion Analyzer b Set the Distortion Analyzer controls as follows FUNCTION Selector SET LEVEL METER RANGE Switch SET LEVEL Range FR...

Page 34: ...er meter indication of 0 9 on the 1 0 scale 5 10 d Adjust the frequency response test set meter control for a set level indication e Change frequency response test set frequency t 20 cps f Readjust the amplitude until the frequency re sponse test set meter indicates set level g The Distortion Analyzer meter shall indicate 0 9 5 NOTE If reading exceeds these limits change A2C25 Typical value for th...

Page 35: ...F THE INSTRUMENT AVOID TOUCHING ANY OF THESE CIR CUITS WITH THE BARE FINGER AS SKIN OILS ARE EXTREMELY CONTAMI NATING IF HANDLING IS NECESSARY WEAR CLEAN COTTON OR RUBBER GLOVES DO NOT USE A PENCIL TO TRACE CIRCUITS IN THE INSTRUMENT GRAPHITE PENCIL LEAD IS AN EX TREMELY GOOD CONDUCTOR AND MN ACCIDENTLY INTRODUCED PATH OF THIS TYPE IS SOMETIMES DIFFICULT TO LOCATE TO AVOID SURFACE CON TAMINATION O...

Page 36: ...TM 11 6625 1576 15 Section V Table 5 8 Model 333A 334A 5 12 ...

Page 37: ...nent lead hole by heating the hole with the iron and inserting a wooden toothpick Re move the toothpick after the solder has cooled and insert the new component lead d To replace components shape new leads and insert them in lead holes Reheat with soldering iron and add a small amount of new solder as required to insure a good electrical connection e Clean excessive flux from the connection and ad...

Page 38: ...TM 11 6625 1576 15 S e c t i o n V T a b l e 5 9 Model 333A 334A 5 14 ...

Page 39: ...Figure 5 10 Troubleshooting Logic 5 15 S e c t i o n V Figure 5 10 T M 1 1 6 6 2 5 1 5 7 6 1 5 ...

Page 40: ...Figure 5 11 Frequency Tuning Assembly 5 16 Model 33A 334A Section V Table 5 10 TM 11 6625 1576 15 ...

Page 41: ...ic diagram An A5 board component location diagram is included wiring data and A5 board component location diagram because the A5 components are not identified on the for the 33A and 334A Distortion Analyzers The sche prlnted circuit boards A partial interconnecting matic diagram illustrate the circuits contained within diagram is also included to assist in reconnecting each assembly as well as the...

Page 42: ...6 2 Figure 6 1 P O Internal Wiring Data Section VI Figure 6 1 Model 333A 334A TM 11 6625 1576 15 ...

Page 43: ...tant General HAROLD K JOHNSON General United States Army Chief of Staff Distribution Active Army USAMB 1 USACDCEC 1 USACDCCEA 1 TOAD 6 USACDCCEA LEAD 8 Ft Huachuca 1 NG None USAR None For explanation of abbreviations used see AR 320 50 U S GOVERNMENT PRINTING OFFICE 1990 0 262 912 30346 SAAD 5 Eighth USA 5 ...

Page 44: ......

Page 45: ......

Page 46: ......

Page 47: ......

Page 48: ......

Page 49: ......

Page 50: ......

Page 51: ......

Page 52: ......

Page 53: ...PIN 016265 000 ...

Reviews: