Models l1049A, l1050A, l1051A Thermal Converters
l1050A
Hewlett-Packard Thermal Converters are true rms indicators,
yielding a dc output voltage proportional to the temperature
rise resulting from the input power. The Models
1l049A,
1l050A
and
1l051A
offer an exceptionally flat response and
nearly constant impedance over a wide frequency range. These
characteristics make the thermal converters ideal to check the
response of precision ac voltmeters, oscilloscopes and amplifiers.
Specifications
Maximum input voltage:
1l049A: 3
V rms;
1l050A:
1
V rms;
1l05lA:
0.45
V rms.
Input impedance:
50f! ±0.15r!
to
10
MHz.
Output voltage for maximum input voltage:
7.5
mV de.
Output impedance:
<
lOr!.
Calibration accuracy
Standard
measurement
Frequency range
In reference to std.
uncertainty
20
Hz to
20
kHz
within
±0.01%
±0.02%
20
kHz to
50
kHz
within
±0.01%
±0.03%
50
kHz to
1
MHz
within
±0.01%
±0.06%
5
Hz to
20
Hz and
within
±0.05%
±0.12%
1
MHz to
10
MHz
10
MHz to
30
MHz
±0.25%
30
MHz to
60
MHz
±0.50%
60
MHz to
100
MHz
± 1.50%
Dimensions:
3"
wide, H4" high,
11/2"
deep
(7,6
x
4,4
x
3,8
cm).
Weight:
net
2.2
oz
(62
g); shipping
1
lb
(450
g).
Price:
HP Model
1l049Ao)(', $125;
HP Model
1l050A*, $125;
HP
11051A'*, $125.
Option
001 *:
calibration to
60
MHz, add
$25.
Option
002*:
calibration to
100
MHz, add
$50.
*
Includes individual calibration report with statement of uncertainty, trace
able to
NBS.
Options include individual correctional data sheet attached to