Hovertech Q2Roller HTR-200 User Manual Download Page 2

User Manual

Air Supply Keypad Functions

Transfer 
Speed 
High

Adjustable/ 

Pressure 

Setting

Transfer 
Speed 
Low

   Standby (Amber LED)

Power Indication

Adjustable Speed

Indication 

(Green Flashing LED) 

Green LEDs

The ADJUSTABLE keypad function has four different settings. 
Each press of the button increases the air pressure and rate 
of inflation. The Green Flashing LED will indicate the inflation 
speed by the number of flashes (i.e. two flashes equals the 
second inflation speed).

STANDBY: Used to stop inflation/air flow (Amber LED indicates 
STANDBY mode).

Instructions for Use

1. Ensure the red deflation valves are capped tightly before beginning

inflation.

2. Place the Q

2

Roller on the bed mattress, and center using the Center

Line Indictor.

3. Tuck the securing apron of the Q

2

Roller under the bed mattress to

keep it in place during use.

4. Remove the backing from the adhesive strips on the Q

2

R Pad, and

center it on top of the Q

2

Roller (if the patient is already on the

mattress, place the device and pad under the patient using a
log-rolling technique).

5. With the patient centered on the device, ensure all bed/stretcher

rails are up.

6. Locate the inflation valve on the OPPOSITE side of the turn direction,

and place the air supply hose over the valve. Press the ADJUSTABLE
button on the HT-Air keypad to initiate air flow. If necessary, press this
button up to 3 more times to increase air flow and rate of inflation.

7. When the turn is complete, remove the hose from the valve and press

the STANDBY button on the air supply keypad to stop air flow. Cham-
ber may not need to be fully inflated to achieve desired position.

8. To quickly deflate the Q

2

Roller, remove the red cap from the deflation

valve. To slowly deflate or adjust the chamber, depress the center of
the black inflation valve.

Product Specifications/Required Accessories

Q2ROLLER

®

 LATERAL TURNING DEVICE

Material:

Polyurethane

Construction:

RF Welded

Width:

Chambers: 46.5" (118 cm) 
(combined chamber width)
Apron: 72" (183cm)

Length:

Chambers: 44.5" (113 cm)
Apron: 44" (112 cm)

LATEX FREE

Model #: HTR-200
Patient Weight Limit: 600 lbs. (272 kg)
Single-Patient Use

Q2R

®

 PAD

Material:

Non-Woven, Breathable PE, Wooden Paper

Construction:

Thermal Bonding

Width:

39” (99 cm)

Length:

39” (99 cm)

T: 1300 106 003

E: [email protected]

www.hmsmedical.com.au

Reviews: