Hotech ADVANCED CT LASER COLLIMATOR User Manual Download Page 10

Advanced CT Laser Collimator Manual (V9.0 =OH=)                                     Page 9 of 20 

a.  This can happen if the telescope was last focused with a diagonal in place (and a diagonal is not installed now), or a focal 

reducer was in the optical train the last time the telescope was focused.  The focal plane is too far out, and is not at the 
normal visual position it would be in if the telescope were focused without a diagonal or focal reducer.  If you do not want 
to use the diagonal during collimation, you will need to adjust the telescope's focus. Continue to 

Step 6.3.b

). 

b.  Adjust the focus to bring at least two laser dots onto the full view on the target plate.  Adjust the focus in one direction first 

to see if any of the laser dots are moving toward the center direction of the target.  If the laser(s) is moving or expanding 
away from the center of the target, reverse the focusing direction to bring at least two laser dots onto the full view of the 
target plate.  Go to 

Step 7

 to collimate your telescope. 

6.4.  Your telescope is grossly out of alignment 

When your telescope is grossly out of alignment, the collimating laser dots may be reflected completely away from the target 
plate.  If you believe this to be the case, begin the collimation process anyway and see if you can bring the three laser dots 
onto the target plate.   
Mechanical Method to get Collimation into Ballpark – Make a preliminary mechanical adjustment of the secondary mirror 
adjustment knobs or screws to achieve a good rough adjustment of the secondary mirror. If you have not replaced the 
secondary mirror adjustment screws with knobs, this would be a good time to do so. Since this procedure is essentially the 
same as the one used to install knobs in place of the screws and greatly expedites secondary mirror adjustment. From this 
point on, the expression “knobs or screws” will be replaced with “knobs”. To make the ballpark adjustment, remove the knobs 
turn counterclockwise one at the time (DO NOT REMOVE THE KNOBS ALL AT ONCE). Mark the knobs with index marks. 
While unscrewing one, gently pull on it while counting the number of turns and fractions of turns until you see exactly how 
many turns and fraction of a turn were made until the threads disengaged. Write down the number of turns and the associated 
knob location (which knob). If you are replacing screws with knobs, do it now. Reinstall each knob – making the noted number 
of turns so it is back to its original position. Do this, one knob at a time. Now, add up and average the number of turns. And re-
adjust all the knobs, one at the time, to the average number of turns. Begin with the knob threaded the most turns into the 
secondary and remove it (CCW).  The secondary mirror will be a bit loose at this point, reinstall the knob to the average 
number of turns. Go to the next longest knob (always one at the time) and remove and re-turning it to the average number of 
turns. Finally set the last knob to the average number of turns. When complete, ensure all the knobs are snug (CW) against the 
secondary mirror. 
Alternate Method for Gross Alignment – A Mirror Reflections Check for a grossly mis-collimated scope can be done by 
observing the reflections of the telescope mirrors by looking directly down the tube. You will have to stop using the collimator. 
Turn off the laser before proceeding. Cut a small hole in the center of a white letter size sheet of paper. Holding the paper at a 
distance of about 7.5 times the diameter of your telescope, look through the hole as straight as you can into the center of your 
Optical Tube Assembly (OTA). Move back and forth until you see white and black (or dark) rings. The white rings are mirror 
reflections of the white paper in the primary and secondary mirrors kind of like barbershop mirror reflections where you sit in a 
chair with a mirror in front of you and a parallel mirror in back of you, and you see reflections of yourself getting smaller and 
smaller off to infinity. The dark rings are the OTA barrel and secondary housing reflections.  Find a distance where you can see 
three white rings. What you want is to see all the rings, both white and dark to be perfectly concentric. If you are reading this, 
the rings are not concentric yet. Now, when you begin to adjust the three secondary mirror collimation knobs always remind 
yourself to work with all rings simultaneously. You can’t get just two rings concentric or you could be off the center optical axis 
one way or the other. To use this method to achieve gross alignment you must work with all the reflected rings (white and dark) 
simultaneously, or the optical axis alignment will be skewed to one side.  

7.0. Collimating the Secondary Mirror 

7.1.  Precautions Before Collimating the Secondary Mirror 

PRECAUTIONS – Be aware of the following when adjusting the secondary mirror collimation knobs (or screws) located on the 

back of the secondary mirror housing assembly: 

a.   Never touch the central screw which holds the secondary mirror (usually only older SCTs will have this central screw). 
b.   The three knobs must be turned in by very small amounts.  No screw being over-tightened or totally unscrewed. When 

secondary mirror collimation adjustment is complete, all three adjustment knobs must be tight. If collimation knobs are 
loose, collimation will not be maintained for long!  ALWAYS finish adjustment by tightening knobs ONLY. In the process of 
adjusting the knobs too achieve collimation – when turning a screw would over tighten it – loosen the opposite knob(s) to 
continue moving the original Knob in the proper direction.  

c.   As you adjust the collimation knobs, observe the collimating lasers three dots position on the collimator target plate to 

determine which knob(s) to turn, and by how much. When collimation is close and the three collimation knobs are snug, 
finishing adjustments will be on the order of 1/32 turn and less. 

NOTE – You will need to constantly re-check telescope & collimator Basic Co-Alignment during the adjustment process to 

make sure proper alignment is maintained.   

 

(Continued Next Page) 

 

Summary of Contents for ADVANCED CT LASER COLLIMATOR

Page 1: ...HOTECH The Synthesis of Innovation and Design User Manual ADVANCED CT LASER COLLIMATOR for CASSEGRAIN TELESCOPES MANUAL VERSION 9 0 WITH OH EDITS...

Page 2: ...Reflector Mirror into your visual back or focuser 6 4 0 Positioning the Laser Collimator at the Proper Distance 6 4 1 Determine the distance between the laser collimator and your telescope 6 5 0 Achie...

Page 3: ...yepiece drawtube or visual back is not square to your primary mirror 11 10 0 Diagonals Reducers and Field Flatteners 11 10 1 Collimation with a Diagonal Accessory Installed 11 10 2 Collimation with a...

Page 4: ...rain type instruments are collimated in similar fashion For telescopes with both adjustable secondary and primary mirrors e g Ritchey Chretien type telescope please email or call us for details All la...

Page 5: ...ir now seen crisp and sharp center the return laser crosshair reflected from the telescope primary mirror in order to adjust the collimator to be perpendicular square to the telescope s optical axis F...

Page 6: ...You will see the projected laser patterns on the wall at various positions Mode 0 Unit off Mode 1 Crosshair laser ON Mode 2 Crosshair laser and three alignment lasers ON Mode 3 Crosshair laser three...

Page 7: ...tance for the telescope 4 1 Determine the distance between the laser collimator and your telescope a Position the collimator one tube length s distance in front of the telescope with the target plate...

Page 8: ...rings then adjust L R until the light beam L R sides are equal in distance to the same target ring d Iterate Step 5 1 c until the donut light beam and collimator target plate ring are as concentric a...

Page 9: ...to diagnose your optics 6 0 How to Read the Diagnostic Result on the Collimator With the telescope and collimator Basic Co Alignment completed Step 5 0 you can determine collimation errors present in...

Page 10: ...to the average number of turns When complete ensure all the knobs are snug CW against the secondary mirror Alternate Method for Gross Alignment A Mirror Reflections Check for a grossly mis collimated...

Page 11: ...erature changes 9 1 Accurate Basic Co Alignment was Not Achieved or is Insufficient It is possible that during collimation Step 7 0 the Basic Co Alignment of the Collimator Telescope Step 5 0 was slig...

Page 12: ...ither or both the focuser or visual back hardware that secures and registers the diagonal or the hardware part of the diagonal that secures and registers the final optic is inadequate the final optic...

Page 13: ...anced Collimator Telescope Z Axes Co Alignment procedures Steps 11 1 c d g h to achieve the best alignment possible within the limits of the telescope s construction Final Optics Axis Squaring Tilt Fo...

Page 14: ...termine the problem if not ideal with a reducer and or field flattener installed refer to Step 10 2 b to choose the best collimation alternative PRECAUTION Collimation and optical alignment beyond the...

Page 15: ...ck or focuser Center the crosshair laser onto the center of the reflector mirror target i If all three crosshairs cannot be perfectly centered then observing the collimator target plate center the two...

Page 16: ...e mechanical system employed to control tilt are presented in Step 11 1 d h Use one of the following mechanical systems to adjust the final optical tilt axis a Accessory Focuser Flange Tilt Setscrews...

Page 17: ...roves tilt alignment to the point that all torque is lost turn it back in CW until the first movement of tilt alignment is observed and stop this is the minimum torque required Then proceed using the...

Page 18: ...Remove the retaining ring and re center the corrector plate to its marked starting position 6 Replace the retaining ring and install all but two of the retainer ring screws to the screw adjustment tor...

Page 19: ...on 6 When the corrector plate is observed to be centered per Step 11 1 g torque CW all the retainer ring screws to the final torque and re check that the corrector plate is centered 7 Re check correct...

Page 20: ...te mechanical measurement is all the old timers had Use the Step 6 4 Mechanical Method to achieve a good rough centering of the Knobs Next remove the Corrector plate perform Step 11 3 a through c ensu...

Page 21: ...Advanced CT Laser Collimator Manual V9 0 OH Page 20 of 20 integrated systems equipment...

Page 22: ...r that is aimed by Tilt adjustment Primary The Telescope Primary Mirror Reflector Mirror forms Reflector Mirror Target Final Optics Mirror Tilt Mirror Tilt Axis Mirror Eyepiece Mirror The Reflector Mi...

Reviews: