Instructions for Use: HotDog

TM

 Head Warming Wrap

INSTRUCTIONS FOR USE

1. Inspect the surface of the Head Warming Wrap for damage (e.g., cuts, holes, loose electrical 

connections). Do not use the Head Warming Wrap if it is damaged.

2. Place the Head Warming Wrap around the patient’s head. 

Warning: DO NOT place the heated area of the Head Warming Wrap under the patient.

3. Insert one end of the electrical cable into the appropriate port on the HotDog Controller. 

Controller Model

Port

WC0X

A

WC5X

A or B

WC71

A

WC77

A, B, C or D

4. Insert the other end of the cable into the electrical connector on the Head Warming Wrap. 

5. Turn the HotDog Controller on and select the desired temperature setting to begin warming. The 

time 

to reach the set-point temperature from 23 C +/-2 C is less than 10 minutes. 

If the Head 

Warming Wrap does not reach the selected temperature within 10 minutes an alarm will sound (Refer 
to the HotDog Controller User Manual.) 

6. If the HotDog Controller alarm sounds when the Head Warming Wrap is connected, do not use the 

Head Warming Wrap until the alarm condition is resolved. (Refer to the “Alarms” section.)

7. When use of the Head Warming Wrap is complete, clean the Head Warming Wrap as necessary. (See 

“Care and Maintenance” section below.)

CARE AND MAINTENANCE

DO NOT continue to use the Head Warming Wrap beyond the labeled expiration date.

DO NOT launder or sterilize as this may damage the Head Warming Wrap.

DO NOT immerse the Head Warming Wrap in liquids.

DO NOT use high-level disinfectants (e.g. gluteraldehyde and peracetic acid) or hydrogen peroxide-based 
solutions to clean the Head Warming Wrap.

DO NOT spray cleaning solutions into the electrical connector.

DO NOT use cleaning or disinfection methods different from those recommended in the Instructions for 
Use without first checking with an authorized service representative to ensure that the proposed methods 
will not damage the equipment.

DO NOT use the Head Warming Wrap if it shows signs of damage or excessive wear such as cuts, holes, 
or loose electrical connectors. Technical staff should perform inspection, such as electrical leakage and 
resistance testing, to determine if it is safe for use. 

DO NOT disassemble the Head Warming Wrap; the product contains no serviceable parts. If service is 
required, call an authorized service representative for assistance.

Storage 

Store the Head Warming Wrap in a dry place and in a manner that prevents it from being cut or crushed.

Summary of Contents for B107

Page 1: ...may result Examples include tissue distal to aortic cross clamping or when vasoconstrictive drugs would lead to severe prolonged vasoconstriction DO NOT warm patients receiving transdermal medication increased drug delivery may occur WARNINGS EXPLOSION HAZARD DO NOT use the HotDog Controller or Head Warming Wrap in the presence of flammable anesthetics or oxygen enriched environments such as hyper...

Page 2: ...multiple warming methods Dispose of the blankets per local regulations The risk of skin irritation caused by pooling of surgical prep solutions under the patient may increase with warming follow surgical prep solution instructions Always use a barrier such as a hair net between the patient and the Head Warming Wrap Head Warming Wrap placement should be considered during X ray imaging as the white ...

Page 3: ... When use of the Head Warming Wrap is complete clean the Head Warming Wrap as necessary See Care and Maintenance section below CARE AND MAINTENANCE DO NOT continue to use the Head Warming Wrap beyond the labeled expiration date DO NOT launder or sterilize as this may damage the Head Warming Wrap DO NOT immerse the Head Warming Wrap in liquids DO NOT use high level disinfectants e g gluteraldehyde ...

Page 4: ...he conductive fabric heaters Dry thoroughly before use Caution DO NOT place the Head Warming Wrap in an autoclave sterilizer automatic washer disinfector or any other high temperature system as this may damage the product Cleaning Disinfection Steps The cleaning steps below are general recommendations and are not meant to replace hospital specific cleaning protocols 1 DO NOT allow cleaning fluids ...

Page 5: ...ngs Medical Device See electronic IFU on website at URL provided EU Authorized Representative IPX2 Protected against dripping water when tilted up to 15 Vertically dripping water shall have no harmful effect when the enclosure is tilted at an angle up to 15 from its normal position The Controller Medical Equipment Classified by Intertek Testing Services NA Inc with respect to electric shock fire a...