7.2
REPLACEMENT PROCEDURES
Technical Service Manual
7 - 3
430-00587-008 (Rev. 2/05)
This section contains safety and equipment precautions, required tools and materials, and
step-by-step procedures for replacing parts in the infusion system. Unless otherwise
stated, always perform the PVT after a replacement procedure.
SAFETY AND EQUIPMENT PRECAUTIONS
Before opening the front enclosure of the infusion system, take all necessary precautions
for working on high-voltage equipment.
CAUTION:
Use proper ESD grounding techniques when handling components.
Wear an antistatic wrist strap and use an ESD-protected workstation. Store PWAs
in antistatic bags before placing them on any surface.
The following tools and materials, or equivalents, are required for the replacement
procedures in this section. In addition, the beginning of each procedure lists tools and
materials required for that specific procedure.
❏
Set of nutdrivers (1/4, 3/16, and 5/16 inch)
❏
Medium size flat blade screwdriver
❏
No. 2 Phillips
®
screwdriver
❏
Fuse puller
❏
Wide-head pliers
❏
Long needle nose pliers
❏
Diagonal cutters
❏
X-acto
®
knife (with square, round, and pointed blades)
❏
Wood chisel, 3/8 inch
❏
Mild solvent (such as isopropyl alcohol)
WARNING:
POSSIBLE EXPLOSION HAZARD EXISTS IF THE INFUSION
SYSTEM IS SERVICED OR REPAIRED IN THE PRESENCE OF
FLAMMABLE ANESTHETICS.
WARNING:
UNLESS OTHERWISE INDICATED, DISCONNECT THE
INFUSION SYSTEM FROM AC POWER BEFORE PERFORMING
ANY REPLACEMENT PROCEDURE.