![Hospira LifeCare XL Series Technical & Service Manual Download Page 124](http://html1.mh-extra.com/html/hospira/lifecare-xl-series/lifecare-xl-series_technical-and-service-manual_2165611124.webp)
Technical Service Manual
8 - 1
430-00587-008 (Rev. 2/05)
This section contains specifications for the 115 VAC Plum XL and 220 VAC LifeCare XL
infusion systems.
The following specifications apply to the Plum XL infusion system only.
PHYSICAL
Dimensions:
Approximately 8.25H x 7.5W x 8.75D inches (including pole
clamp)
Weight:
Approximately 7.75 lbs (with battery)
Casing:
High-impact plastic
ELECTRICAL
Power Requirements:
100-130 VAC, 50 to 60 Hz, less than 35 W
Power Cord:
Hospital-grade AC (mains) cord. 10 feet long, with
transparent plug
Fuses:
0.5 A, 250 V, slo-blo
Battery:
Sealed lead-acid, rechargeable 8 V battery, internal to
infusion system. Accessible for ease of field replacement,
with color-coded leads and polarized connector
Battery Operation
(XL/XLM):
A fully charged new battery provides eight hours of operation
at 125 mL/hr, or 1000 mL total volume delivered, whichever
occurs first
Battery Operation (XLM
with DataPort):
With DataPort circuitry disabled, a fully charged new battery
provides eight hours of operation at 125 mL/hr, or 1000 mL
total volume delivered, whichever occurs first. With
DataPort enabled, a fully charged new battery provides
approximately five hours of operation at 125 mL/hr, or 625
mL total volume delivered, whichever occurs first
Note:
Do not connect DataPort when infusing
Recharge:
The battery charges whenever the infusion system is
connected to AC (mains) power. If the infusion system is
operating at 125 mL/hr, a full recharge takes approximately
16 hours. If the infusion system is turned to OFF/CHARGE,
recharge takes approximately eight hours