16
Status description
It explains the screen display at each monitoring status.
Normal
Screen in a normal status is as follows.
Fig. 7 Normal screen
Oil film warning (OIL WARNING)
The oil film warning is triggered when the detection area which laser light reflection level is
over the set detection level exceeds the detection area setting (Warning), and when the
continuous judgment time passes.
For setting the judgment criteria, refer to “ Detection level setting ” (page 35), “ Detection area
setting ” (page 36), and “ Continuous judgment time setting ” (page 37).
The screen when the oil film warning occurs is as follows.
Fig. 8 Oil film warning screen
[NORMAL RUN] is displayed.
Current detection level is colored in
the unit of 1 mA.
Current detection area is colored in
the unit of 10
.
[OIL WARNING] is displayed.
Current detection level is colored in
the unit of 1 mA.
Current detection area is colored in
the unit of 10
.
Contact output release button is dis-
played.