Changing the automatic power off setting (default: OFF)
D-74
61
●
Changing the automatic power off setting (default: OFF)
You can set the instrument to automatically turn OFF when there is no key operation for
a certain period of time.
This function is disabled during automatic data memory saving or remote operation using
an external device.
1. Press the
key to enter the setting mode.
2. Press the
keys to select the "MAINT" (maintenance setting) and then press
the
key.
3. Press the
keys to select the "AT OFF" (automatic power off setting) and
then press the
key.
4. Select "ON" to use this function, or select "OFF" not to use it. And then press
the
key.
When "ON" is selected, enter the automatic power off time and press the
key.
The setting range is 1min to 30 min.
To return to the setting mode, press the
key.
key
key
key
key
Selected "ON"
and press the
key