![horiba LAQUA COND DS-72 Instruction Manual Download Page 84](http://html1.mh-extra.com/html/horiba/laqua-cond-ds-72/laqua-cond-ds-72_instruction-manual_2162237084.webp)
Chapter 8 Periodic Inspection Mode
72
HORIBA
Span check result
Criteria: ±0.5% ±1digit of the full scale
±1.5% ±1digit of the full scale only when 19.00 S/m is entered.
Linearity check result
Criteria: ±0.5% ±1digit of the full scale
±1.5% ±1digit of the full scale only when 10.00 S/m is entered.
Temperature check result
Indication error for each entry (regulated value ±0.4
C).
8.2 Comment Input
A comment can be entered up to 100 characters. Use this function to record periodical
checks, etc.
Tap
to use the function.
To delete the content input previously, tap
.
Summary of Contents for LAQUA COND DS-72