ISFET pH Sensor

0141

 

Instruction Manual

Thank you for purchasing a HORIBA ISFET pH electrode.
ISFET pH sensor (0141) is a component of ISFET pH
electrode (0040-10D). Read this manual with that of ISFET
pH electrode carefully before use.

1

Cautions on use

Safety Precautions

Points of concern

The materials used of the electrode may dissolve in the
sample depending on the specified sample or lengthy
measurement. Do not eat the whole, or the part around
where the electrode sensor touched, of measured
samples.
If measured sample gets on the skin, wash off it com-
pletely with soap and water.

Do not touch the sensing part and sensor terminal, or It
may be damaged by static electricity.

Store the product in dry conditions. Storage area should
be kept away from high temperature, high humidity, and
direct sunlight. 

Do not wet the electrode connector and the sensor ter-
minal.

Avoid continuous use in high temperature, acids or alkali
sample of 0.01 mol/L or thicker. Failure to comply may
cause decline life period of the sensor.

White powder or solution may be generated at the liquid
junction. This is leaked internal solution of the reference
electrode, and not a symptom of trouble. Rinse the
sensor with ion-exchange water before use.

If  internal solution of the reference electrode gets on the
skin, wash off it completely. In case that the solution gets
into the eyes, continue to flush the eyes with running
water. And after washing the eyes thoroughly, get
medical help immediately.

2

Specifications and Components

2.1

Packaged Contents

2.2

Components

CAUTION

Handle carefully the sharp electrode tip to not get hurt. 

If calibration standard solution gets on the skin, wash off it 
completely. In case that the solution gets into the eyes, continue to 
flush the eyes with running water for at least 15 minutes. And after 
washing the eyes thoroughly, get medical help immediately.

Name

Q’ty

Sensor (0141)

1 pc

Instruction manual (this book)

1 booklet

Sensor terminal

End cap

O-ring

Sensing part
/Liquid junction

2.3

Specifications

3

Preparation

3.1

Assembling sensor and converter

Note

Do not touch the sensing part and sensor terminal, or It may
be damaged by static electricity.

When tightening the nut, no gap should be left between the
nut and the sensor. If the nut is not tightened enough, water
may enter the inside of the sensor and the sensor may be
damaged.

1. Remove the end cap from the sensor, and the seal from

the liquid junction. Make sure that the O-ring is attached
to the sensor.

2. Turn the nut to the left to remove it from the converter.

3. Insert the sensor to the converter slowly checking that

the O-ring is untwisted, and match the 

 marks on the

sensor and converter. Then turn the nut to the right to fix
the sensor.

Sensor model

0141

Storage temperature 
range

0

C to 40

C

Length

100 mm

Sensing device

ISFET (Ion-sensitive field-effect transistor) 

Internal electrode

Silver/ silver chloride electrode

Reference electrode 
internal solution

Carboxymethyl cellulose (CMC), glycerin, 
saturated KCl 

Liquid junction material

Porous sintered polyethylene

Sensor material

Tantalum pentoxide, LCP, epoxy resin, 
polyethylene resin 

Height of sensing part

1 mm (from the tip

 

Applicable electrode

0040-10D

Sensor terminal

End cap

O-ring

Liquid junction seal

Nut

 mark

Nut