22
7.5
Initialization mode
All settings and all calibration data are reset to the factory default values.
7.6
Software version display mode
The current software version is displayed.
This information may be asked for depending on your inquiry.
1.
Press and hold the MEAS switch for over 3
seconds in the measurement mode to
enter the special setting mode.
All items appear on the LCD, and then the dis-
play changes as shown right.
2.
Press the CAL switch until Init appears.
3.
Press the MEAS switch for 0.5 seconds.
appears.
4.
Press the CAL switch for over 2 seconds.
All settings and all calibration data are reset to
the factory default values.
When initialization is completed, End and
appear.
5.
Press the ON/OFF switch to turn OFF and
ON again.
1.
Press and hold the MEAS switch for over 3
seconds in the measurement mode to
enter the special setting mode.
All items appear on the LCD, and then the dis-
play changes as shown right.
2.
Press the CAL switch until Init appears.
3.
Press the CAL switch for 0.5 seconds.
The software version is displayed.