52
Has the internal
solution been
replaced?
Replace the internal solution
before each measurement.
If not, the internal solution
concentration will change and it
will cause measurement errors.
(
Is there a sufficient
amount of internal
solution?
Make sure that the surface of the
solution is just below the internal
solution fill port.
Was the standard
solution made using
pure water?
Prepare the standard solution
using pure water. Preparing the
standard solution with water other
than pure water will cause
measurement errors.
(
―
Is calibration
performed before
each measurement?
Perform calibration before each
measurement.
(
(page 19))
―
Is the electrode
temperature same
as measurement
place?
Wait until the temperature of the
electrode becomes the
temperature of that place before
using the electrode.
―
Is the ion selective
membrane or liquid
junction dirty?
Rinse the electrode. A dirty
electrode will cause measurement
errors or a decrease in the
response rate during calibration.
(
Has the internal
solution
concentration
changed?
Replace the internal solution
completely.
(
―
Check item
Solution
View in video
Summary of Contents for 6560S-10C
Page 1: ...Instruction Manual Combination type Chloride Ion Selective Electrode 6560S 10C ELECTRODE ...
Page 8: ......
Page 20: ...8 ...
Page 56: ...44 ...
Page 66: ...54 ...
Page 80: ...68 ...
Page 83: ...71 Material External dimensions Unit mm 16 5 O D 18 85 150 ...