3. Outside Interference Factors affecting SpO
2
measuring accuracy
•
Intravascular dyes such as indocyanine green or
methylene blue
•
Exposure to excessive illumination, such as
surgical lamps, bilirubin lamps, fluorescent lights,
infrared heating lamps, or direct sunlight.
•
Insufficient coverage of the y probe can interfere
with readings. Make sure to have the Y-Probe
centered in the middle of the Hopkins Wrap.
•
Vascular dyes or externally used products such as
nail enamel or tinted skin care
•
Excessive patient movement
•
Placement of a sensor on an extremity with a
blood pressure cuff, arterial catheter, or
intravascular line
•
Exposure to the chamber with high pressure
oxygen
•
There is an arterial occlusion proximal to the
sensor