SPECIFICATIONS
19
Table 4
Guidance and manufacture’s declaration – electromagnetic immunity – for ME EQUIPMENT and
ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
The 526018 is intended for use in the electromagnetic environment specified below.
The customer or the user should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m 80 MHz
to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the 526018, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Portable and mobile RF communications
equipment should be used no closer to any
part of the including cables, than the recom-
mended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.167√Ρ
d = 1.167√Ρ 80 MHz to 800 MHz
d = 2.333√Ρ 800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a
should be less than the compliance level in
each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol: