Rectal Use:
Lubricate the probe cover with a water-soluble jelly for easier
insertion; do not use petroleum jelly, as it may cause an inaccurate reading.
Gently insert the tip of the probe no more than 1/2 inch into the rectum.
STOP
if you meet any resistance. The sensing unit is on the very tip of the probe
and there is no need to insert the probe deep into the rectum.
Dispose of the used probe cover and wash the thermometer as recommended.
The peak temperature should be reached in approximately 25-35 seconds.
Axillary Use:
Wipe armpit with a dry towel. Place the probe in the
patient’s armpit (Figure 3) and keep the patient’s arm pressed
firmly against the body. In order to achieve a better axillary reading,
a minimum measuring time of 90 seconds or longer is
recommended, depending on placement of the probe and
closure of the arm against the body.
Battery Replacement
1. When the symbol “ ” appears, the battery needs to
be replaced.
2. Remove battery compartment lid (Figure 4).
3. Use a pointed object such as a toothpick to pry loose the
old battery. Avoid using any sharp metal object.
4. Place new 1.55V coin-type battery (LR41 or equivalent) in the chamber with the
positive (+) side facing up or towards the display of the unit.
5. Replace the battery compartment lid.
Cleaning and Disinfecting
Do not use corrosive detergent to clean the unit.
Clean and disinfect the thermometer before and after each use.
Cleaning Process
1. Wipe probe with a cloth dipped in a disinfectant solution such as 70%
isopropyl (rubbing) alcohol or soap and water.
2. Rinse off disinfectant residue.
Care and Storage
• Do NOT drop or crush; this device is not shock proof.
• Do NOT dismantle or modify device other than replacing the battery.
• Do NOT store the device in an extreme environment with direct sunlight or high
temperatures.
• Store the unit in the protective case while not in use.
LIMITED WARRANTY
This thermometer is guaranteed for a period of 24 months from date of purchase. This warranty
does not cover batteries, or damage resulting from improper handling or use. Specific legal rights
stipulated in your state may guide this warranty. If you have a legitimate claim under this warranty,
send item(s) postage paid to Service Center, Attn: Repair Dept., 6850 Southbelt Dr., Caledonia, MI
49316. Please include your name and address, phone no., proof of purchase, and a brief note ex-
plaining the problem, your return address and $2.00 to cover shipping and handling.
IB p/n 93-410HOP-00 rev 6 (4/25/2020)
Printed in China
LATEX
Figure 4
Specifications
Size:
5”L x 3/4”W x 3/8”H (12.7cm L x 1.9 cm W x .95 cm H)
Weight:
0.388 oz. (11 grams)
Display:
LCD, Liquid Crystal Display
Battery:
1.55V (LR41)
Battery Life:
Estimated 4,500 measurements
Measurement Time:
Oral 30* seconds / Rectal 25-35 seconds / Axillary 90 seconds
Measurement Range:
90.0°F to 109.9°F (32.0°C to 42.9°C)
96.0°F to 107.0 °F
±
0.2°F / 1°C (34.0°C to 42.0°C)
at a room temperature of 64.4°F to 82.4°F (18.0°C to 28.0°C)
Accuracy:
90.0°F to 96.0°F
±
0.4°F / 0.2°C (32.0°C to 34.0°C)
and 107.0°F to 109.9°F (42.0°C to 42.9°C)
Measurement Mode:
Real measurement with peak-hold method
Audible Alarm:
1 beep at start-up. Continuous beep at measurement completion
Auto Turn-off:
In approximately 8 ±1 minutes
Memory:
Last reading
Operating Conditions:
50°F to 104°F (10°C to 40°C)
Humidity of 15% to 95% (non-condensing)
Storage Conditions:
-13°F to 131°F (-25°C to 55°C)
Humidity of 15% to 95% (non-condensing)
Standards:
Complies with EN 12470-3, ASTM E 1112-00,
IEC 60601-1, IEC 60601-1- 2, IEC 60601-1-11
Electromagnetic compatibility: device fulfills the stipulations
of standard IEC 60601-1-2
The manufacturer recommends a calibration check every 2 years.
*30 second response time achieved in water bath test.
BATTERY
Service Center
6850 Southbelt Dr.
Caledonia, MI 49316
Inspected in the U.S.A.
Made in China
Read the instruction manual carefully before using
this device, especially the safety instructions, and
keep the instruction manual for future use.
Type BF applied part
Batteries and electronic devices must be disposed of
in accordance with the locally applicable
regulations, not with domestic waste.
Figure 3