Operating Instructions
Integrated Transducer UVA-AS102
U326_UVAintAS102_B_e_070717
6/13
www.hoentzsch.com
A3.2.2 Power supply
Before connecting please check that the power supply is within the specification. Remove the cover of the
integrated UVA to reveal the type plate with the relevant information.
A3.2.3 Analog output v
Remove the cover of the integrated UVA to reveal the type plate with the relevant information regarding
analog output.
The terminal value of the analog output can be parameterised with the PC software UCOM via the RS232
interface. The customer-specific programmed value can be found in the accompanying documents.
P
= protective earth
4-20 mA
Resistance max. 400 Ohm
0-10 V
Impedance 1 kOhm