Dr. K. Hönle Medizintechnik GmbH 

 

idromed

®

 5 PS

 

 

 

G

EBRAUCHSANWEISUNG

 

 

3

 

 

1. 

I

MPORTANT INFORMATION ON IDROMED

®

 

5

 

PS 

 

Symbol for important safety instructions. 
Please follow these instructions carefully to prevent health risks and damage to 
property. 

In case of the following conditions the idromed

®

 5 should not be used or should only 

be used after consulting your doctor: 

 

cardiac arrhythmia 

 

electronically controlled implanted devices (e.g. pacemaker) 

 

metallic implanted devices in the area where the current flows 

 

metallic intrauterine implanted devices (loop) 

 

pregnancy 

 

bigger skin diseases 

 

insensibility for pain 

Prolonged usage of high frequency surgery units on a patient can cause burns under  
electrodes. 

Application of the electrodes close to the ribcage can increase the risk of ventricular 
fibrillation. 

Effective  current  densities  over  2  mA/cm²  at  all  electrode  faces  require  higher 
attention from the user. 

 

During treatment the electric circuit needs to be completed constantly, i.e. maintain 
at least one hand or foot in each tub. 

Conductive  material  (e.g.  metal,  water  pipes  etc.)  must  not  be  touched  during 
treatment. 

All jewellery on hands and feet must be removed before starting treatment. 

Any damage to skin (minor injuries, scratches etc.) on the palms of the hands, soles 
of the feet, cuticles or at the armpits must be coated with vaseline or a fatty cream 
before treatment, as current permeability will be higher than normal on those spots. 
Please do only cover these spots! 

Current  intensity  is  set  according  to  individual  requirements  of  the  user.  The  ideal 
current  intensity  evokes  a  pleasant,  tingling  sensation  (not  on  the  treatment  of  the 
armpits).  The  values  listed  in  these  instructions  are  the  maximum  values  allowed. 
Current  intensity  chosen  by  the  user,  however,  should  normally  be  much  lower, 
otherwise  there  is  a  danger  of  burning  the  skin  (e.g.  the  usual  range  for  the 
treatment of the armpits is about 1 – 2 mA). 

Moisten  the  sponges  from  inside  and  outside,  until  they  don’t  drip  (do  not  wring 
them out). If the sponges are too dry, there is a danger of burns! 

Attention,  there  is  a  danger  of  burns,  when  the  metallic  electrodes  are  in  direct 
contact with the skin. When treating armpits, the sponge electrodes have to be kept 
still and with constant pressure under the armpits, otherwise fluctuations of current 
will occur on the display of the unit. 

 

Never short-circuit the metallic electrodes! 

Summary of Contents for 110001

Page 1: ...1 OPERATING INSTRUCTIONS idromed 5 PS Art No 110001 Art No 110003 Before putting the unit into operation and starting treatment read this operating manual carefully 110001 3 S0408 e ...

Page 2: ...arge treatment tubs art no 995130 2 x distance grids for large treatment tubs art no 995131 Content 1 Important information on idromed 5 PS 3 1 1 Symbols and warning on the unit 4 1 2 Electromagnetical compatibility 4 2 Application of the idromed 5 PS 5 2 1 Fields of Application 5 2 2 Conditions of Application 5 2 3 Preparation of the Treatment 5 2 4 Treatment 6 2 4 1 Treatment of hyperhidrosis on...

Page 3: ...d during treatment All jewellery on hands and feet must be removed before starting treatment Any damage to skin minor injuries scratches etc on the palms of the hands soles of the feet cuticles or at the armpits must be coated with vaseline or a fatty cream before treatment as current permeability will be higher than normal on those spots Please do only cover these spots Current intensity is set a...

Page 4: ...cology Unit contains electronic parts and may not be disposed in the household rubbish 1 2 ELECTROMAGNETICAL COMPATIBILITY idromed 5 PS uses high frequency energy exclusively for its internal function Therefore its HF emission is very low and it is unlikely that surrounding electronic devices are disturbed HF emissions acc to CISPR 11 group 1 idromed 5 PS is determined for the use in all facilitie...

Page 5: ...e manufacturer b For treatment you should separate the two parts of the case To do this open the case completely put it on a plain surface and separate the 2 parts carefully in opposite directions see arrows on the hinges on the case c Fill the 2 treatment tubs with tap water to a depth of approx 3 4 cm finger and toe joints should be just covered Please see also the warnings on page 10 point 4 3 ...

Page 6: ...reatment of the feet or of hands and feet 25 mA For the treatment of the armpits 5 mA If the current intensity selected is too high you may experience an unpleasant tingling sensation in the extremities treated during the session If this occurs a lower current intensity has to be selected otherwise there is a danger of burns 2 4 1 TREATMENT OF HYPERHIDROSIS ON THE HANDS AND OR FEET Self test Maxim...

Page 7: ...would break the circuit by taking one hand out of the tubs While adjusting the current hands or feet may be taken out of the treatment tubs and while the unit is still running without any danger Please dry your hands before regulating the current at the unit Regulate the current by turning the knob right or left and close the circuit again by putting again your hands in the treatment tubs Repeat t...

Page 8: ...ect the electrodes with the positive and the negative pole of the unit Self test Maximum armpits 5mA Values hands 15mA feet 25mA Switch on the unit while pushing down the turning knob Pay attention that the turning knob is on switch setting The unit carries out a self test This action takes some seconds Set the therapy current Place your hands feet or armpit sponges 2mA Adjust the required current...

Page 9: ...ns the remaining treatment time At the end of the treatment remove the sponge electrodes from the armpits Turn the knob to the switch setting and switch off the unit by pushing down the knob Always check the position of the turning knob Position immediately after switching on the unit and the self test completes 3 IMPORTANT TIPS AND INFORMATION To simplify therapy the current intensity in mA for y...

Page 10: ...please contact Customer Services 4 3 Current intensity is adequate Check the connections of the plug and electrodes and the function of the unit Please check if the water you are using has been treated with a water softener By using softened water the conductivity is reduced In this case please use untreated tap water or use non carbonated still mineral or table water For the treatment of the armp...

Page 11: ...lly the general Terms of Sales and Delivery of the company Dr K Hönle Medizintechnik GmbH shall apply These shall be submitted to the operator upon the signing of the contract at the very latest The warranty for the idromed 5 PS is 24 months Warranty and liability claims are excluded in case of personal injury or damage of property if these are attributable to one or more of the following causes T...

Page 12: ...l be alight Once the unit has been switched off the lamp will be out Protection class II all insulated MDD classification IIa Protection against intrusion of liquids IPX1 CE Mark Output wave form Manufacturer A L T Lichttherapietechnik GmbH Thura Mark 10 D 06780 Zörbig Exclusive sales Dr K Hönle Medizintechnik GmbH Viktor Frankl Str 20 D 86916 Kaufering Tel 0049 8191 30555 0 Fax 0049 8191 30555 50...

Reviews: