Configurable Parameter Reference Guide
MAN0984_Iss 4_01/19
Touchpoint Plus
Pt. No. 3011M5001
136
Technical Handbook
Chapter 17.
To avoid confusion between letter O and number 0, Ø represents number zero or logic null.
Parameter Name
Parameter
Description
Applies
to
Read/
Write
Range
Default Value
Remark
Channel Enable
Channel Enable/Disable
mA Input R/W
Disable/Enable
Disable
Channel Number
Number
mA Input R/W
1~16
-
An unique number
(Auto assigning 1~16 LED )
Channel ID will be
automatically assigned
Channel Tag
25 characters
mA Input R/W
Alphanumeric
Character
Range
50 chars
Detector Name
20 characters
mA Input R/W
Alphanumeric
Character
Range
40 chars
Write available when:
Detector Name- User
Detector
Target Gas Name
16 characters
mA Input R/W
Alphanumeric
Character
Range
16 chars
Write available when:
Detector Name- User
Detector
Target Gas Unit
%, ppm, %LEL, %VOL, mA, User (6chars)
mA Input R/W
Alphanumeric
Character
Range
%
Write available when:
Detector Name- User
Detector
mg/m3, g/m3, %VOL,
ppm, %LEL, UEG,
Ratio, %LEL x M, ppm x m,
EG x m, %Vol x m, LEL.m,
LEL%.m, %V/V, mA, kppm,
oC, oF, %RH
Target Gas Digit
Numbers of decimal point
mA Input
0~3
1