SPM Single Point Monitor
SPM Technical Handbook
4-7
4.8 Output Terminal Status
Alarm Relay Terminals
Condition
Alarm Level
1
Alarm Level
2
*Instrument
Fault
N/O
N/C
N/O
N/C
N/O
N/C
Power Off
O
C
O
C
O
C
Power On, No Gas
O
C
O
C
C
O
Power On, Gas (Level 1)
C
O
O
C
C
O
Power On, Gas (Level 2)
C
O
C
O
C
O
Power On, Fault
O
C
O
C
O
C
Power On, Alarm Test
C
O
C
O
C
O
Power On, Verify
O
C
O
C
C
O
Power On, Tape Load
O
C
O
C
C
O
Power On, No Gas, Relay Disable
O
C
O
C
O
C
Power On, Gas, Relay Disable
O
C
O
C
O
C
Power On, Fault, Relay Disable
O
C
O
C
O
C
Power On, Alarm Test, Relay Disable
O
C
O
C
O
C
Power On, Verify, Relay Disable
O
C
O
C
O
C
Power On, Tape Load, Relay Disable
O
C
O
C
O
C
Summary of Contents for SPM
Page 8: ...SPM Single Point Monitor Midas Technical Handbook 1 1 1 Operation ...
Page 15: ...SPM Single Point Monitor SPM Technical Handbook 2 1 2 Installation ...
Page 20: ...SPM Single Point Monitor Midas Technical Handbook 3 1 3 Maintenance ...
Page 24: ...SPM Single Point Monitor Midas Technical Handbook 4 1 4 Specifications ...
Page 36: ...SPM Single Point Monitor SPM Technical Handbook 5 1 5 Options ...
Page 53: ...SPM Single Point Monitor SPM Technical Handbook A 1 A Bi Directional Communications Protocol ...
Page 61: ...SPM Single Point Monitor SPM Technical Handbook B 1 B Warranty Statement ...