SIGNALPOINT PRO OPERATING MANUAL MAN0853 Issue 3 10/08
24
9. When the gas reading is stabilised, exit the test mode by swiping the magnet
once over the oval mark during the test mode. The instrument then displays “
rESEt “ for 15 seconds.
10. Wait for the instrument to time out of the “ rESEt “ display mode and the
instrument then enters the calibration mode.
11. Recalibrate the detector (see step 3 of section 8.3.1 for toxic version or step 3
of section 8.3.2 for Oxygen version).
Sensor Cover
Sensor retaining ring
Transmitter
Diagram 7: Replacing plug in sensor
Sensor