background image

RDD100 User Guide 

 

6                                                                                                                                                             www.honeywell.com

 

OPERATION 

 

3.1 

Start Up Messages 

When the RDD100 is powered up, it goes through a test of its display segments and its error LED.  The error LED 
is located beneath the least significant digit of the display.  All segments should light for a brief period and then 
the display transitions to “rdd100”, “Sr 2.0”, and “rEAdY”.  Note: older RDD100 units may not display “Sr 2.0”.  
Newer RDD100 units may also have additional LED in the lower right hand corner of the display which flashes as 
each packet of data is received. 
 
If a transducer is properly set up and connected to the RDD100, pressure readings will appear on the display. 

 

3.2 

Error Indications 

3.2.1 

No Transducer Connected 

If the RDD100 does not detect a transducer, the display will go through the segments and LED check, display 
“rdd100”, display “Sr 2.0”, and finally display “rEAdY” before going dark with only the error LED illuminated.  Note: 
older RDD100 units may not display “Sr 2.0”. 

 
3.2.2 

Incorrect Baud Rate 

If the RDD100 sees a transducer, set to the wrong baud rate, the display will go through the segments and LED 
check, 

display “rdd100”, display “Sr 2.0”, display “rEAdY” and then hold with the “rEAdY” message.  Note: older 

RDD100 units may not display “Sr 2.0”. 

 
3.2.3 

Not Set Up To Continuously Transmit 

If the RDD100 sees a transducer that has not been set up to continuously transmit, the display will go through the 
segments and LED check, 

display “rdd100”, display “Sr 2.0” and finally display “rEAdY” before going dark with 

only the error LED illuminated.  Note: older RDD100 units may not display “Sr 2.0”. 

 
3.2.4 

Wrong Polarity Of DC Power  

The RDD100 is equipped with the reverse polarity protection diodes at its power input.  The instrument will 
operate with either polarity. 

 
3.2.5 

Transducer Disconnected While Transmitting 

If the transducer is disconnected while transmitting, the last reading received will remain on the display and the 
error LED will light, indicating that no further updates are being made.  Reestablishing connection to the 
transducer will result in a continuation of reading updates and the error LED will turn off. 
NOTE: the RDD100 may appear to be in this condition if the transducer has been programmed with a long 
integration time or idle count yielding long durations between readings.  If changed from its factory default 
settings, it is recommended that the connected transducer be programmed to provide 1 or more readings per 
second to prevent this possible confusion.  (The typical factory default reading rate is 5/second.) 

 
3.2.6 

Over/Under-Ranged Transducer Indications

 

In the event that a transducer, experiencing an applied pressure or temperature which is over or under the 
specified range, is connected to the RDD100, the LED display will flash on-off-on-

off…with the current readings to 

indicate the over/under-ranged condition. 
 
Over/under-ranged con

ditions on older RDD100’s (prior to Sr 2.0) will indicate the following: 

 

Sequence #1: 

If a transducer that is in an over/under ranged condition is connected to the RDD100 and then the power is 
applied to the RDD100, the result will be a start-up message t

hat continuously displays “rEAdY”. 

 

Sequence #2: 

If the RDD100 is powered up and then an over/under ranged transducer is connected to the RDD100, the result 
will be a blank display and the error LED will light.   
Once the over/under range condition is removed, normal operation of the RDD100 will be restored. 

 

 
 

Summary of Contents for RDD100

Page 1: ...Honeywell Pressure Sensor Products RDD100 Remote Digital Display User Manual ...

Page 2: ...ed Parameters 3 Operation 3 1 Start Up Messages 3 2 Error Indications 3 2 1 Transducer Not Connected 3 2 2 Incorrect Baud Rate 3 2 3 Not Set Up To Continuously Transmit 3 2 4 Wrong Polarity Of DC Power 3 2 5 Transducer Disconnected While Transmitting 3 2 6 Over Under Ranged Transducer Indications 4 Power Requirements 4 1 Voltage 4 2 Current 4 3 Power Connection 5 Operating Environment 5 1 Operatin...

Page 3: ...dependent upon the pressure display units selected Although many of the supported transducers also have a 0 5V analog output only the digital output is used by the RDD100 as it provides better resolution and accuracy Any of 16 different pressure display units may be selected for use with the RDD100 A list of available pressure units is shown in section 2 4 below A more complete description of the ...

Page 4: ...te for the transducer is shown below The commands below all use a broadcast address 99 to allow them to operate correctly even if the transducer address has been changed from the factory default setting which is address 01 With the PC and transducer operating at the baud rate found to return a non garbled start up message Type 99we cr Carriage Return Resp 99WE Command Echo Type 99bp n9600 cr Resp ...

Page 5: ...was in the original psi scale For example when a 10 psia device is switched to DU LCOM the full scale digital pressure output will be 60 000 instead of 10 000 Also a 500 psi device will output full scale 600 00 instead of 500 00 PFS generally provides the highest readout resolution available when using the RDD100 readout 2 5 Continuous Transmit Mode Type 99we cr Resp 99WE Command Echo Type 99mo p2...

Page 6: ... Power The RDD100 is equipped with the reverse polarity protection diodes at its power input The instrument will operate with either polarity 3 2 5 Transducer Disconnected While Transmitting If the transducer is disconnected while transmitting the last reading received will remain on the display and the error LED will light indicating that no further updates are being made Reestablishing connectio...

Page 7: ...ed 2 Separate cables are available for transducers with Metal or Plastic electrical connectors Standard cable length is 6 feet 1 8 m Connectors may be purchased from Honeywell to construct custom length cables Due to RS 232 limitations recommended maximum length is 50 ft 15 m 3 12 VDC plug is 2 1 x 5 5 x 10 mm I D x O D x length Center pin may be either positive or negative polarity RDD100 Wall mo...

Page 8: ...RDD100 User Guide 8 www honeywell com Honeywell International 12001 State Highway 55 Plymouth MN 55441 Tel 1 800 323 8295 www honeywellmps com ADS 14124 December 2012 ...

Reviews: