GASLAB Q2 AFTER DELIVERY AND AT THE PLACE OF USE
Information for general use
Rev. M / 73023639
77
GasLab Q2
Digital outputs
There are four digital outputs (electrically isolated passive output circuits) in the connection box
(TB6-terminals, details see figure 5.6).
The output at TB6-5 is a
breaker (NC, “normally closed”), the other outputs are makers (NO,
“normally open”). The maximum load per channel is 28.8 V DC/120 mA. The maximum pulse rate is
25 Hz.
Cable type refer to
5.4.2 Cables (power supply / communication)
the maximum length is 250m
≙
820ft. The figure also shows the appropriate terminals in the connection box. The wiring in the
drawing is only designed to illustrate the screening using an example.
Figure 5.10: Example connection of digital outputs
In addition to the wiring, various enSuite settings are required before this interface
can be used.
Changing the IP address basic network and I/O settings