Configuration
3.2 Configuration Set Up Procedure
ATTENTION:
In Table 3-1, under the
Press
column:
Hold
means to hold the button down until the display changes.
Momentarily
means to press and release the indicated button.
From the Online display, follow this procedure.
ATTENTION:
If no key is pressed for 60 seconds, the display will abort the entry mode and default to
Online Display.
Table 3-1 Configuration Set Up Procedure
Step
Operation
Press
Display
1
Enter
Cell
Constant
Selection
(if applicable)
MODE
Hold
CnSt
(
for 1 second
) then,
(C
urrent Cell Constant Selection
)
Edit Cell Constant
MODE
Hold
Flashing Display – You are now in EDIT
mode
(Cell Constant Value)
Select desired Cell
Constant
Momentarily
To select 0.01, 0.1(default), 1, 10, 25, 50
Save the Cell
Constant
MODE
Momentarily
Selection for Cell Constant
2
Enter
PV Type
Selection
MODE
Momentarily
PtYP
(
for 1 second
) then,
(Current Selection)
Edit PV type
Selection
MODE
Hold
Flashing Display – You are now in EDIT
mode
(V
alue of current PV Type selection
)
Select desired PV
Type
(selection
determined by cell
constant)
Momentarily
To select:
S, mS (default), rES, PPb, PPm, PPt,
Conc
Save the PV Type
continued
MODE
Momentarily
Selection for PV Type
14
DirectLine® DL423 Sensor Module User Manual
9/11