Daily Operation
6. When
appears, close the cylinder
valve.
7. When the calibration is successfully completed,
appears. If it fails,
appears with the
flashing non-compliance symbol
.
8.
appears with the number of remaining
days until the next calibration.
9. Remove the calibration cap from the detector.
If repeated attempts continue to fail, contact Honeywell Analytics or your
distributor for technical assistance.
2.10 Calibration Cap
The calibration cap, which is provided as part of the package, is required
to do a bump test or calibration. Use this method to apply calibration gas
to the gas detector.
1. Connect a tube from a cylinder of cali-
bration gas to the calibration cap’s right
spigot.
2. To attach the calibration cap to the gas
detector, hook the cap’s left clip to the
corresponding groove of the detector
and snap the right clip into place.