Calibration Handbook
28
VA301M / S301M
4. Adjusting the Span
Turn on the regulator and inject span gas at the specified flow rate.
The span gas is now flowing into the unit.
Wait 2 minutes before starting the Set Span calibration.
Press the Enter key, and using the Scroll key enter the VA password.
Using the Scroll key select Zone 1 (built-in CO sensor) or Zone 2 (remote sensor) and press Enter.
Scroll to get the Set Span menu and press Enter to go in.
Using the Scroll and Enter key, input the concentration of span gas that’s going to be used.
Wait will be displayed on the screen until calibration is completed.
Scroll until Back is displayed, press Enter.
Scroll until Quit is displayed, press Enter to exit.
S301M Calibration Procedure
1. Warm-Up
Period
Turn the unit on for a minimum of fifteen (15) minutes.
2. Connectiing the Hardware
Screw the regulator to the appropriate calibration gas cylinder.
Adjust the regulator with a flow rate of according to specification.
S301M CALIBRATION INFORMATION
Gas
Scale
Warm-up
Flow Rate
CO
0 - 255 ppm
15 min
0.5L/min
NO2
0 - 10.2 ppm
15 min
0.5L/min
C3H8
0 - 25.5 % Vol
15 min
100 ml/min
Combustibles
0 - 102 %
15 min
100 ml/min
Connect the regulator outlet to the sensor as shown on.
or
58 Liter
Calibration
Gas Cylinder
103 Liter
Calibration
Gas Cylinder
Calibration cap
S301M Calibration Equipment
3. Remote Calibrating the S301M Sensor
When using the S301M in a network of sensors, connected to a VA301C controller (with firmware 3.0 or
Summary of Contents for 705 Series
Page 1: ...HA University Honeywell Fixed Gas Detection Calibration Handbook ...
Page 2: ......
Page 6: ......
Page 8: ......
Page 9: ...Calibration Guide VA201T VA201M ...
Page 13: ...Calibration Guide VA301D2 GasPoint II ...
Page 14: ......
Page 18: ......
Page 19: ...Calibration Guide 301IRFS S301D2 with VA301EM ...
Page 20: ......
Page 25: ...Calibration Guide VA301M S301M ...
Page 26: ......
Page 30: ......
Page 31: ...Calibration Guide VA301W Wireless Transmitter ...
Page 32: ......
Page 35: ...Calibration Guide 705 Sensepoint Combustible ...
Page 36: ......
Page 38: ......
Page 39: ...Calibration Guide 90DM4 ...
Page 40: ......
Page 43: ...Calibration Guide Apex Transmitter and Sensor ...
Page 44: ......
Page 49: ...Calibration Guide E3 Point ...
Page 50: ......
Page 53: ...Calibration Guide Manning EC F2 ...
Page 54: ......
Page 57: ...Calibration Guide Manning EC F3 O2 ...
Page 58: ......
Page 61: ...Calibration Guide Manning EC F9 NH3 ...
Page 62: ......
Page 65: ...Calibration Guide Manning EC P2 ...
Page 66: ......
Page 69: ...Calibration Guide Searchline Excel ...
Page 70: ......
Page 83: ...Calibration Guide IAQPoint ...
Page 84: ......
Page 87: ...Calibration Guide IAQPoint2 ...
Page 88: ......
Page 92: ......
Page 93: ...Calibration Guide IR F9 AirScan iR ...
Page 94: ......
Page 97: ...Calibration Guide Midas Gas Detector ...
Page 98: ......
Page 104: ......
Page 105: ...Calibration Guide Searchpoint Optima Plus ...
Page 106: ......
Page 113: ...Calibration Guide Series 3000 XPIS ...
Page 114: ......
Page 119: ...Calibration Guide Sensepoint Toxic ...
Page 120: ......
Page 124: ......
Page 125: ...Calibration Guide SQN8x ...
Page 126: ......
Page 133: ...Calibration Guide Sensepoint XCD ...
Page 134: ......
Page 139: ...Calibration Guide Sensepoint XCD RTD ...
Page 140: ......
Page 145: ...Calibration Guide XNX Universal Transmitter ...
Page 146: ......
Page 154: ......
Page 155: ...Calibration Guide Calibration Kits and Equipment ...
Page 156: ......
Page 162: ......
Page 163: ......