10 |
GB
GB
| 11
Table 1 For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration-electromagnetic emissions
The Ear Thermometer is intended for use in the electromagnetic environment
specified below. The customer or the user of the Ear Thermometer should assure that
it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
guidance
RF emissions CISPR 11
CISPR 11
The Ear Thermometer uses RF energy
only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The Ear Thermometer is suitable
for use in all establishments other
that domestic and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations
Flicker emissions
IEC 61000-3-3
Not applicable