24
VER: A001 202203
B.
With the cuff wrapped around your upper arm, and place the
finger of the opposite hand (index finger recommended) gently
on the Advanced IHB detection area, then make sure the finger
place at the correct position, press
button to start
measurement. All display units appear on the screen
for 1.5
seconds.
*Note
!
Do not inflate the cuff until it is wrapped around your upper arm.
Please clean the finger before taking the Advanced IHB Detection
measurement.
Advanced IHB detection will have error result if blood pressure measurement
is applied on the same arm as the finger detected.
C.
After all symbols disappear, the display will
show “00”. The monitor is “Ready to
Measure” and will automatically inflate to
the level that is right for you.
D.
As the cuff inflates, the monitor automatically determines your
ideal inflation level. This monitor detects your blood pressure
and pulse rate during inflation. The Pulse Symbol (
) flashes at
each pulse beat. Remain still and do not move until the entire
Measurement Procedure
*Note!
When press the
key, the measurement of blood pressure and
Advanced IHB detection will be both activated, if only Advanced IHB
detection feature will be measured, please press the
key again to stop
the measurement of blood pressure.