-27-
SYMBOLS/SYMPTOMS
CONDITIONS/CAUSES
INDICATION/CORRECTION
Unit does not turn on when
START/STOP
button is
pushed.
Worn-out batteries.
Replace them with 2 new AAA
(LR03) alkaline batteries.
Battery polarities have been
positioned incorrectly.
Re-insert the batteries in the
correct positions.
Measuring Error Symbol
appears when blood
pressure value displayed is
excessively low or high.
Cuff has been placed
incorrectly.
Wrap the cuff properly so that it is
positioned correctly.
Measuring Error Symbol
Air circuit abnormality. Cuff
tube may not be plugged into
monitor correctly.
Check cuff connection. Measure
again.
Measuring Error Symbol
Inflation pressure exceeding
300 mmHg.
Switch the unit off, then measure
again.
Measuring Error Symbol
Can’t determine blood
pressure measurement data.
Wrap the cuff properly and keep
steady. Measure again.
Cuff is worn improperly
Wrap the cuff snugly so that it is
positioned correctly.
If you have any question about the
cuff wearing and/or measurement
result, please consult your
healthcare professional.
Measuring Error Symbol
If the device cannot be
connected to paired
Bluetooth-enabled device over
45 seconds, LCD will display
Error message “E4” and
Bluetooth will be turned off.
Please press
button for 3
seconds to start Bluetooth function.
Excessive Body Motion
Detector
Notice:
The measured
blood pressure reading may
not be accurate if the icon is
displayed.
Body movement during
measurement, especially, the
movement on the wrist the
blood pressure monitor is worn
on.
e.g. Talking, moving or
shaking of the arm with the
cuff on while measurement.
Measure again. Keep steady
during measurement.
Cuff is worn improperly, or the
shape of the wrist is unusual,
excessive gap might be exist
between the cuff and the wrist.
Wrap the cuff properly and keep
steady. Measure again.
If you have any question about the
cuff wearing and/or measurement
result, please consult your
healthcare professional.
Troubleshooting