7.2.3 Assigning a Static IP Address
The most efficient way to assign the instrument a static IP address is to use the Console in the
Holzworth GUI. The Console can be used to send the commands from Appendix B which are used
to change the instrument from DHCP to Static, set the static IP,
etc.
Users must first establish a
USB connection or a direct Ethernet connection as referenced in section
7.1
and
7.2
, respectively.
Once a connection has been established, launch the Console
.
Now users can begin sending the
ASCII commands from Appendix B. The commands should be sent in the order shown in the list
and the Console screenshot below.
1.
Send the command to change the static IP address.
:IP:ADDR:<value>
2.
Send the command to change the subnet address.
:IP:SUBNET:<value>
3.
Send the command to change the gateway if necessary.
:IP:GATEWAY:<value>
4.
Send the command to change from DHCP to Static.
:IP:STATUS:STATIC
5.
Power cycle the instrument when prompted.
When the instrument fully powers back on (5-10 second power up) it will come up with the static IP
settings and can be connected to the LAN.