TruNode
®
Gamma Probe System User Manual
27
Immunity to RF Wireless Communications Equipment
Test
Frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Max
Power
(W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
385
380 –
390
TETRA 400
Pulse
modulation
b)
18 Hz
1.8
0.3
27
450
430 –
470
GMRS 460,
FRS 460
FM
c)
± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704 –
787
LTE Band 13,
17
Pulse
modulation
b)
217 Hz
0.2
0.3
9
745
780
810
800 –
960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0.3
28
870
930
1720
1 700 –
1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1,
3,
4, 25; UMTS
Pulse
modulation
b)
217 Hz
2
0.3
28
1845
1970
2450
2 400 –
2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2
0.3
28
5240
5 100 –
5 800
WLAN
802.11
a/n
Pulse
modulation
b)
217 Hz
0.2
0.3
9
5500
5785
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because
while it does not represent actual modulation, it would be worst case.