Guidelines and manufacturer's statement - electromagnetic compatibility
EN
11.5
Guidelines and Manufacturer's Statement - Electromagnetic In-
terference Immunity
Note 1: The higher frequency range applies for 80 and 800 MHz.
Note 2: These guidelines are probably not realizable in all cases. The distribution
and spread of electromagnetic quantities differs depending on the absorption
and reflection of buildings, objects, and people.
a
The field strength of stationary transmitters such as base stations of wireless
phones and cell phones, ham radio operators, AM and FM radio and TV stations
can theoretically not always determined in advance. A study of the installation
site should be considered to determine the electromagnetic environment con-
cerning the stationary transmitter. If the measured field strength at the pro-
posed Aquilex Fluid Control System installation and operation site exceeds the
concordance levels listed above, the Aquilex Fluid Control System should be mon-
itored to document proper functionality and operation as intended. If unusual
performance characteristics are observed, additional measures may be required
such as changing orientation or the location of the Aquilex Fluid Control System.
Electromagnetic
interference
immunity tests
Test level
Compliance
Electromagnetic environ-
ment guidelines
Conducted HF
interference
quantities
according to IEC
61000-4-6
Radiated HF
interference
quantities
according to IEC
61000-4-3
3 V
eff
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
In compliance
In compliance
Portable and mobile wire-
less devices should not be
used in closer proximity to
the Aquilex Fluid Control
System (including cables/
lines) than the recom-
mended safety distance
calculated based on the
transmitting frequency
and the applicable for-
mula. Recommended
safety distance:
d = 1.2
P for 150 KHz
to 80 MHz
d = 1.2
P for 80 MHz to
800 MHz
d = 2.3
P for 800 MHz
to 2.5 GHz
With
P
as the rated output
of the transmitter in watts
[W] according to the infor-
mation provided by the
manufacturer of the trans-
mitter and
d
as recom-
mended safety distance in
meters [m].
The field strength of sta-
tionary transmitters for all
frequencies tested on site
a
should be lower than the
concordance level.
b
Interference is possible in
the proximity of devices
featuring the following
pictograph.
Summary of Contents for AQL-100
Page 6: ......
Page 50: ...46 Appendix EN 15 Appendix 15 1 Test Log Date Tests Performed Results Comment Signature ...
Page 52: ......
Page 54: ......
Page 102: ......
Page 104: ......
Page 152: ......
Page 154: ......
Page 202: ......
Page 204: ......
Page 252: ......
Page 254: ......
Page 302: ......