Annual Inspection
29
EN
8.4
Flow Rate Test
Performing flow rate test
The test set-up is depicted in Figure 8-1, Flow rate test.
1. Turn the system on. (See 5.4, Turning On the Aquilex System)
2. Insert the tube set into pump and close bag clamps.
3. Hang the fluid bags onto the hooks of the bag holder.
4. Spike bags and open bag clamps.
5. Insert hysteroscope tube into measuring container.
6. Set intrauterine pump pressure to 150 mmHg.
7. Press the
Prime
button.
8. The roller wheel starts to turn to purge air from tubing and complete auto-
matic lumen calibration.
9. Once automatic lumen calibration finishes (
~
20 sec), press
Pause/Resume
button.
10. Empty measuring container.
11. Re-Insert hysteroscope tube into measuring container.
12. Press the
Pause/Resume
button.
13. After one minute, press the
Pause/Resume
button. The measuring container
should contain approximately 800 ml of fluid.
14. The acceptable tolerance is ±25 ml/min.
Record results in the test log in Section 15.1. Test is successful if results fall within
acceptable tolerance limits.
Figure 8-1 Flow rate test
Summary of Contents for AQL-100
Page 6: ......
Page 50: ...46 Appendix EN 15 Appendix 15 1 Test Log Date Tests Performed Results Comment Signature ...
Page 52: ......
Page 54: ......
Page 102: ......
Page 104: ......
Page 152: ......
Page 154: ......
Page 202: ......
Page 204: ......
Page 252: ......
Page 254: ......
Page 302: ......