System Operation
21
EN
5.9
Inserting the Inflow Tube Set
(See Figure 5-9, Tube set elements) The inflow tube set consists of three tube sec-
tions, a Y-connector
(4)
and two bag spikes
(2)
. The three tube sections are: roll-
er wheel section
(6)
, inflow section
(5)
, and hysteroscope section
(8)
.The bag
spikes
(2)
are used to connect the tube sections to the bags.
The Luer lock connector
(9)
connects the hysteroscope tube with the hystero-
scope.
Open outer packaging
1. Inflow tube set - To be carried out by non-sterile nurse:
Open outer packaging of the inflow tube set.
A sterile nurse then removes the inner tube set package and opens it.
Connect to hysteroscope
2. To be carried out by sterile nurse:
Keep the blue Luer lock connector
(9)
in the sterile area and hand the tube
end with the bag spikes
(1)
to the non-sterile nurse.
Figure 5-9 Tube set elements
(1)
Protective caps
(2)
Bag spikes
(3)
Tubing clamps
(4)
Y-connector
(5)
Inflow section
(6)
Roller wheel section
(7)
Pressure chamber with membrane
and RFID transponder
(8)
Hysteroscope section
(9)
Luer lock connector (blue)
(10)
Roller wheel connector
(7)
(5)
(8)
(1)
(2)
(4)
(3)
(9)
(6)
(10)
Figure 5-10 Inserting the tube set
(1)
Bag spikes
(2)
Fluid bags
(3)
Bag clamps
(5)
Inflow tube
(6)
Roller wheel tube
(7)
Pressure chamber with mem-
brane and RFID transponder
(8)
Hysteroscope tube
(9)
Luer lock connector (blue)
1
2
3
(9)
(8)
(3)
(2)
(5)
(7)
(1)
(6)
Summary of Contents for AQL-100
Page 6: ......
Page 50: ...46 Appendix EN 15 Appendix 15 1 Test Log Date Tests Performed Results Comment Signature ...
Page 52: ......
Page 54: ......
Page 102: ......
Page 104: ......
Page 152: ......
Page 154: ......
Page 202: ......
Page 204: ......
Page 252: ......
Page 254: ......
Page 302: ......