background image

 

 

User Manual 

 

Version 1.0                                                                                                                                  6 
 

 

II.

 

Safety          

  

 

IMPORTANT Safety Items to be Observed. 

 

Safety Warnings

 

! DANGER (may cause serious damage to property and or casualties) 

1.  Please carefully read this User Manual prior to operating the instrument and observe the 

instructions on safe operation. 

2.  Ensure that only trained staff operate this instrument  
3.  The power source must be grounded reliably and away from any sources of 

electromagnetic interference. 

4.  Ensure that the instrument and its accessories are free of any potential defects. 
5.  Confirm that the voltage and frequency of the power supply matches the specified voltage 

and frequency specified prior to use. 

6.  Ensure the cover of the junction box is properly installed prior to operation. 
7.  Always wear applicable PPE whilst operating the instrument. 
8.  Place the instrument on a stable, dry, clean, antiskid, and fireproof surface. 
9.  Ensure there is sufficient ventilation gaps and spacing above and around the instrument 
10. Ensure the instrument is stable and level and will not move due to vibration upon start-up. 
11. Liquid must not enter the casing 

 

! WARNING (may cause property damage or personal injury) 
 

12. The working environment must be free of any flammable, explosive, volatile or corrosive 

substances.  

13. 

Only accessories listed as ‘optional accessories’ for use, can guarantee operational safety. 

14. Ensure handling and disposal of packaging is compliant to local environmental laws and 

regulations. 

 
     

! ATTENTION (may affect operational performance or service life) 

15. The instrument is deenergised only when the power cable is disconnected from the power 

source. 

16. Ensure the instrument and its accessories are protected external vibration. 
17. The  Company  reserves  the  right  to  modify  the  design  and  technical  data  of  the  pump 

without notice.  

18. Only accredited and qualified professional repair technicians can open the instrument or 

conduct required repairs. Persons performing repairs on the instrument other than those 
selected or approved by the Company shall operate to void any warranty contained 
hereinabove for the product 

19. Do NOT operate this instrument immediately after transport, wait 1-2 days before 

operating. The instrument must be used in accordance with this User Manual. 

Summary of Contents for 10303001

Page 1: ...User Manual Version 1 0 1 10303001 10303002 10303003 Aquilon Chiller User Manual...

Page 2: ...ort 7 III Aquilon Introduction 8 Overview 8 IV Technical Specifications 9 V Proper Use 10 VI Inspection 10 VII Operating Instructions 11 Preparation and Precautions Prior to Use 11 7 1 User Interface...

Page 3: ...ns certain representations of the potential dangers from the instrument or equipment Buyer is solely responsible for conducting any research necessary to learn the hazards involved for any of its inte...

Page 4: ...trical power surges vi improper storage and handling of the Product or vii use of the Product in combination with equipment or software not supplied by the Company If the Company determines that Produ...

Page 5: ...BE LIABLE FOR INDIRECT INCIDENTAL CONSEQUENTIAL SPECIAL OR PUNITIVE DAMAGES INCLUDING LOSS OF BUSINESS OR PROFITS UNDER ANY THEORY OF LIABILITY HEREUNDER For claims under the warranty please contact y...

Page 6: ...not move due to vibration upon start up 11 Liquid must not enter the casing WARNING may cause property damage or personal injury 12 The working environment must be free of any flammable explosive vol...

Page 7: ...ent or equipment or Service without charge at its sole discretion Buyer assumes sole responsibility for any reliance on or use of such assistance and information and the Company makes no warranty ther...

Page 8: ...They are designed for continuous reliable operation When the liquid is not in use the liquid should be discharged Included are alarm functions such as timing and over temperature protection A Pt100 p...

Page 9: ...Cooling capacity at 10 W 1200 1500 2500 Refrigerant R404A R410A Electrical performance Power Consumption W 800 1300 1600 Sensor PT100 Controller Display Type Liquid Crystal Safety protection Compresso...

Page 10: ...rer VI Inspection Unpack the equipment carefully and check for any damage which may have arisen during transport In the event of identified damage please contact serviceusa hollandgreenscience com The...

Page 11: ...the instrument casing e The liquid must cover the evaporator f Use an appropriate liquid for desired temperature range g When the liquid is water pay close attention to the temperature set point to p...

Page 12: ...nd operation stops To set a timer press the SETTINGS key When the TIME value flashes enter the desired time using they arrow keys Leave at zero if a timer is not needed Silence Indicator Indicates a s...

Page 13: ...to flash Set the desired temperature value with or keys The button allows for movement between the digits Confirm the set value with another press of button Time Settings Press the button The last va...

Page 14: ...t Stop Cooling Operation Before the refrigeration is turned on please press key to let the liquid circulate in the pipeline in advance After setting the operating parameters press button for 4 to 5 se...

Page 15: ...ent should be cleaned only by a clean soft rag and detergent The power must be disconnected before any maintenance or cleaning Do not clean the instrument with any corrosive cleaning solutions If the...

Reviews: