Inlet Nozzles

For isokinetic microbiological air samplings in external air flows, e.g. in air
ducts, inlet nozzles are available to adapt the nominal flow rate of the sam-
pling head  to the one of the sample air. 

The flow rate of the sample air has to be measured with a flow meter to
determine the required inlet nozzle from the table above.

The  cleaned  inlet  nozzle  is  mounted  on  the  inlet  pipe  of  the  transverse
adapter. The sealing is provided by an O-ring on the shaft of the inlet noz-
zle.  When  using 

MBASS30

as  air-conveying  device,  make  sure  that  the

outgoing air is flowing in the same direction as the sample air.

3

Applications

The transverse adapter is an optional accessory for the air sampling
heads 

LKS 30

and 

LKS100

and allows to draw in the sample air hor-

izontally. Additionally, isokinetic microbiological air samplings and
the  loading  of  PU  foam  cylinders  for  chemical  air  samplings  are
possible.  The  transverse  adapter  is  used  instead  of  the  top  part  of
the air sampling head.

Components

The transverse adapter consists of the adapter head and the inlet pipe. The
inlet pipe gets screwed into the adapter head and can be unmounted for
cleaning. Only screw on the transverse adapter hand-tight. The sealing is
provided by a gasket on the pipe.

Loading of PU Foam Cylinders

PU foam cylinders with a diameter of 50 mm and a length of 50 to
70  mm  can  be  loaded  for  chemical  air  samplings.  Up  to  two  PU
foam cylinders can be inserted behind each other in the pipe. When
operating with 

MBASS30

, 300 liters of sample air should be drawn prior

to the sampling without a PU foam cylinder for acclimation. The sampling
procedure is described in the relevant guidelines (e.g. DIN ISO 16000 parts
12 and 13).

Remark when using the air sampling head 

LKS100

:

The air sampling head 

LKS100

has a much lower air

resistance  than  the 

LKS  30 

when  operating  at

30 l/min to load o PU foam cylinder. Depending on
the resistance of the PU foam cylinder, this may lead
to the error message ”No sampler active”. By placing
the included 55 mm circular filter onto the jet plate,
the air resistance of 

LKS100

can be adjusted.

Remarks when using 

MBASS30

as air-conveying device:

The usage of a binderless glass fiber filter is not possible due to its high flow
resistance. Manual checking of the air flow before and after the sampling,
as required by some guidelines, is not possible.

Inlet Nozzle

No.

Flow with

LKS 30

Flow with

LKS100

Article No.

1

-

3 m/s

02-171

2

1,5 m/s

5 m/s

02-172

3

2 m/s

7 m/s

02-173

4

3 m/s

10 m/s

02-174

5

5 m/s

16 m/s

02-175

6

7 m/s

-

02-176

Sample air flow

Sampling in external air flow with transverse adapter and inlet nozzle

Transverse Adapter and Inlet Nozzles

Operating Manual for Transverse Adapter for Air Sampling Heads

2

The Transverse Adapter

Transverse Adapter and Inlet Nozzles

Operating Manual for Transverse Adapter for Air Sampling Heads

Isokinetic Microbiological Air Samplings