64
Functions of the Device
I-Trigger
(only in modes S, T and ST)
The I-Trigger determines the trigger sensitivity for
detecting inspiration. It can be set to three levels.
1
10 ml
2
15 ml
3
25 ml
Go to the Clinic-Menu to set the I-Trig-
ger. Then select
I-Trigger
by using the keys
and
. Press the
key. Now the device will
prompt you to enter the PIN. Now select the
desired level by pressing the keys
and
. Press
To confirm. Next the parameter E-Trigger is set.
E-Trigger
(only in modes S and ST)
The E-Trigger determines the trigger sensitivity for
detecting expiration. It can be set to three levels.
1
60 % of Peak
fl
ow
2
70 % of Peak Flow
3
80 % of Peak
fl
ow
Press the
key. Now select the desired level by
using the keys
and
. Press
to con
fi
rm.