- 20 -
8.13 CHECk THE ALARMS
You must adhere to the procedure given in chapter “Functional Check” of the User's Manual.
8.14 DATA BACkuP
If it is proposed to change the set parameters during a SRC or maintenance, the contents of the
device memory must be read and saved. If necessary, the data is to be analyzed. no maintenance
is necessary to the PC software.
8.15 STAnDARD VALuES THE FACTORy SETTIng
Parameter values in accordance with the factory setting (through a PC or direct on the device)
general Parameters
Mode
PCV
Display I:E
I:E
Display Pressure Units
hPa
Display Screen
STD
Safety Lock
Manual
Safety Lock (Machine)
OFF
Volume
50 %
Pressure Deviation
3.0 hPa
Max. Inspiration time
4 s
Leakage
OFF
Max. Oxygen
OFF
Min. Oxygen
OFF
PIn code
1302
Device Language
Eng
Parameter PCv
IPAP
20.0 hPa
PEEP
5.0 hPa
Frequency
12 bpm
I:E
1 : 1.5
Chute
3
Inspiration trigger
OFF
Min. Volume
OFF
Additional pressure
5.0 hPa (not active, Min. volume is set to OFF)
Max. Volume
1.00 l
Min. Volume
0.20 l
Max. Frequency
30 bpm
Summary of Contents for CARAT I
Page 24: ... 24 10 Exploded Views 10 1 Exploded View C Inner Casing CARAT I ...
Page 25: ... 25 10 2 Exploded View C Inner Casing CARAT II ...
Page 26: ... 26 10 3 Exploded View C Inner Casing CARAT II ...
Page 27: ... 27 10 4 Exploded View A Top Cover CARAT I ...
Page 28: ... 28 10 5 Exploded View A Top Cover CARAT II ...
Page 29: ... 29 10 6 Exploded View E Sensor Manifold CARAT I ...
Page 30: ... 30 10 7 Exploded View E Sensor Manifold CARAT II ...
Page 31: ... 31 10 8 Exploded View B Lower Cover CARAT rear view ...
Page 32: ... 32 10 9 Exploded View B Lower Cover CARAT front view ...
Page 33: ... 33 10 10 Exploded View Assembling CARAT I ...
Page 34: ... 34 10 11 Exploded View Assembling CARAT I ...
Page 35: ... 35 10 12 Exploded View Assembling CARAT II ...