15. Technical results
EMC requirements
IEC 60601-1-2 Edition 4.0 (2014)
Environment of intended uses:
Home Healthcare Facility Environment
Summary of Test Results
Test
Standard
Class/ Severity level
Test
result
Emission
(IEC 60601-1-2 section 7)
Radiated emission
Freq. range: 30 - 1000
MHz
CISPR 11
Group 1 Class B
Complies
Immunity
(IEC 60601-1-2 section 8)
Immunity from
Electrostatic discharge
(ESD)
IEC 61000-4-2 8 kV contact discharges &
15 kV air discharges
Complies
Immunity from radiated
electromagnetic fields
IEC 61000-4-3 3.0; 10.0 V/m; 80 MHz 2.7
GHz,
80% AM, 1 kHz
Complies
Immunity from Proximity
field from wireless
communications
equipment
IEC 61000-4-3 List of frequencies,
from 9 V/m up to 28 V/m,
PM (18 Hz or 217 Hz), FM 1 kHz
Complies
Immunity from power
frequency magnetic field
IEC 61000-4-8 30 A/m @ 50 Hz
Complies