background image

“Sensor reading within limits” if you want the alarm to 
clear once the sensor reading returns to the normal 
range between the high and low alarm limits. Select 
“Cleared with button press” if you want the alarm to 
remain on until you press the Alarm/Stats button on the 
logger. 

9.

 

Click OK to save alarm settings. 

Notes: 

 

Once the logger is launched, alarms will trip as 
determined by these settings. Logger alarms will display 
on the LCD screen. Note that the alarm limits are only 
checked when the logger’s LCD screen refreshes every 15 
seconds. 

 

The actual values for the high and low alarm limits are set 
to the closest value supported by the logger. For 
example, the closest value to 85°F that the UX100 series 
logger can record is 84.990°F and the closest value to 
32°F is 32.043°F. In addition, alarms can trip or clear 
when the sensor reading is within the logger 
specifications of 0.02°C resolution. This means the value 
that triggers the alarm may differ slightly than the value 
entered. For example, if the High Alarm is set to 
75.999°F, the alarm can trip when the sensor reading is 
75.994°F (which is within the 0.02°C resolution). 

 

When you read out the logger, high and low alarm levels 
will be displayed on the plot along with “Chan <#> Alarm 
Tripped” and “Chan <#> Alarm Cleared” events showing 
when the sensor alarm tripped and cleared. The “Chan 
<#> Alarm Cleared” event contains the value that was 
furthest out of range for the sensor before the alarm 
cleared (see the Points table for the actual value). 

Burst Logging 

Burst logging is a logging mode that allows you to set up more 
frequent logging when a specified condition is met. For 
example, let’s say the logger is recording data at a 5-minute 
logging interval and burst logging is configured to log every 10 
seconds when the temperature goes above 85°F (the high limit) 
or falls below 32°F (the low limit). This means the logger will 
record data every 5 minutes as long as the temperature 
remains between 85°F and 32°F. Once the temperature reaches 
90°F, for example, the logger will switch to the faster logging 
rate and record data every 10 seconds until the temperature 
falls back below the high limit (or 85°F in this case). At that 
time, logging then resumes every 5 minutes at the normal 
logging interval. Similarly, if the temperature falls to 30°F, for 
example, then the logger would switch to burst logging mode 
again and record data every 10 seconds. Once the temperature 
rises back to 32°F, the logger will then return to normal mode, 
logging every 5 minutes. 

To set up burst logging: 

1.

 

Select Burst for Logging Mode in the Launch Logger 
window. If Burst has already been configured for this 
logger, click the Edit button in the Launch Logger 
window. Note that you cannot set up alarms with burst 
logging. You also cannot select the Stop Logging option 
“Never (wrap when full)”” if burst logging is configured. 

2.

 

In the Burst Logging window, select a sensor. In the 
example at right, the temperature sensor was selected. 

3.

 

Select the High Limit checkbox if you want to set up a 
condition in which burst logging will occur when the 
sensor reading rises above the high limit value. Type in 
the value or drag the red upper slider (see example 
below). 

4.

 

Select the Low Limit checkbox if you want to set up a 
condition in which burst logging will occur when the 
sensor reading falls below the low limit value. Type in the 
value or drag the blue lower slider. 

 

5.

 

Repeat steps 2 through 4 for the other sensor, if desired. 

6.

 

Set the burst logging interval, which must be less than 
the logging interval. Select either a preset burst logging 
interval or select Custom and enter your own interval. 
Keep in mind that the more frequent the burst logging 
rate, the greater the impact on battery life and the 
shorter the logging duration. 

7.

 

Click OK when done. This will return you to the Launch 
Logger window. Click the Edit button next to Logging 
Mode in the Launch Logger window to make additional 
changes. 

Notes: 

 

Once the logger is launched, the high and low burst 
logging limits are only checked when the logger’s 
LCD screen refreshes once every 15 seconds. Therefore, 
if you set the logging interval to less than 15 seconds and 
the sensor reading falls outside the limits, the burst 
logging will not begin until the next 15-second refresh 
cycle. 

 

If high and/or low limits have been configured for more 
than one sensor, then burst logging will begin when any 
high or low condition goes out of range. Burst logging will 
not end until all conditions on all sensors are back within 
normal range. 

 

The actual values for the burst logging limits are set to 
the closest value supported by the logger. For example, 
the closest value to 85°F that the logger can record is 
84.990°F and the closest value to 32°F is 32.043°F. 

 

Burst logging mode can begin or end when the sensor 
reading is within the logger specifications of 0.02°C 
resolution. This means the value that triggers burst 

Summary of Contents for UX100-023

Page 1: ...al Cable Length 1 83 m 6 ft Logger Logger Operating Range Logging 20 to 70 C 4 to 158 F 0 to 95 RH non condensing Launch Readout 0 to 50 C 32 to 122 F per USB specification Logging Rate 1 second to 18 hours 12 minutes 15 seconds Logging Modes Fixed interval normal burst or statistics Memory Modes Wrap when full or stop when full Start Modes Immediate push button date time or next interval Stop Mod...

Page 2: ...ns of each symbol in the table below LCD Symbol Description The logger is waiting to be launched Press and hold the Start Stop button for 3 seconds to launch the logger The logger has been launched with a push button stop enabled press and hold the Start Stop button for 3 seconds to stop the logger Note If you also launched the logger with a push button start this symbol will not appear on the dis...

Page 3: ...hen remain on for 10 minutes The LCD screen refreshes every 15 seconds regardless of the logging interval selected in HOBOware If you choose a logging interval less than 15 seconds the data will be recorded at the faster interval but the sensor readings will only be updated on the screen every 15 seconds When the logger has stopped logging the LCD screen will remain on until the logger is offloade...

Page 4: ...kbox the LCD will not show the current readings status or other information while the logger is logging You will however be able to temporarily turn the LCD screen on by pressing the Start Stop button for 1 second if you select this option 10 Click the Start button to launch the logger Disconnect the logger from the computer and deploy it using the mounting materials see Mounting the Logger After ...

Page 5: ...logging mode again and record data every 10 seconds Once the temperature rises back to 32 F the logger will then return to normal mode logging every 5 minutes To set up burst logging 1 Select Burst for Logging Mode in the Launch Logger window If Burst has already been configured for this logger click the Edit button in the Launch Logger window Note that you cannot set up alarms with burst logging ...

Page 6: ...deviation series four for temperature and four for RH with values calculated and logged every 5 minutes based on the 30 second sampling To set up statistics 1 Select Statistics for Logging Mode in the Launch Logger window If Statistics has already been configured for this logger click the Edit button in the Launch Logger window 2 Click the Maximum Minimum Average and Standard Deviation checkboxes ...

Page 7: ...B 2 and follow these instructions 1 Remove the tape fastening the sensor cap to the receptacle Discard the tape 2 Grasp the cap and pull firmly to remove it Discard the old cap 3 Note the orientation of the small circuit board containing the RH sensor Pull it out and discard 4 Push gently but firmly to install the new sensor in the same orientation 5 Push the RH new sensor cap onto the receptacle ...

Page 8: ...n fire Do not expose the contents of the battery to water Dispose of the battery according to local regulations for lithium batteries HOBOware provides the option of recording the current battery voltage at each logging interval which is disabled by default Recording battery life at each logging interval takes up memory and therefore reduces logging duration It is recommended you only record batte...

Reviews: