HOBO
®
U10 Temp/RH Data Logger (Part # U10-003)
© 2005–2012 Onset Computer Corporation
Part #: MAN-U10-003, Doc #: 11196-B
Inside this package:
•
HOBO U10 Temp/RH
Logger
•
Mounting kit with magnet, hook-and-loop
tape, and 3/8” double-sided tape.
Doc # 11196-B, MAN-U10-003
Onset Computer Corporation
Thank you for purchasing a HOBO data
logger. With proper care, it will give you
years of accurate and reliable measurements.
The HOBO U10-003 is a two-channel
Temperature/Relative Humidity Data Logger
with 10-bit resolution and capacity for
52,000 measurements. The logger uses a
direct USB interface for launching and data
readout by a computer.
HOBOware
®
software is required for logger
operation. Visit www.onsetcomp.com for
details.
Specifications
Measurement
range
Temperature: -20° to 70°C (-4° to 158°F)
RH: 25% to 95% RH
Accuracy
Temperature: ± 0.53°C from 0° to 50°C (± 0.95°F
from 32° to 122°F), see Plot A
RH: ± 3.5% from 25% to 85% over the range of
15° to 45°C (59° to 113°F), see Plot B; ± 5%
from 25% to 95% over the range of 5° to 55°C
(41° to 131°F), see Plot B
Resolution
Temperature: 0.14°C at 25°C (0.25°F at 77°F),
see Plot A
RH: 0.07% @ 25°C and 30% RH
Drift
Temperature: 0.1°C/year (0.2°F/year)
RH: <1% per year typical
Response time in
airflow of 1 m/s
(2.2 mph)
Temperature: 10 minutes, typical to 90%
RH: 6 minutes, typical to 90%
Time accuracy
Approximately ± 1 minute per month at 25°C
(77°F), see Plot C
Operating range
Logging: -20° to 70°C (-4° to 158°F); 0 to 95%
RH (non-condensing)
Battery life
1 year typical use
Memory
64K bytes (52,000 10-bit measurements)
Weight
26 g (0.82 oz)
Dimensions
45 x 60 x 20 mm (1.8 x 2.38 x 0.77 inches)
The CE Marking identifies this product as
complying with the relevant directives in the
European Union (EU).
Plot A
Plot B
Plot C
Connecting the logger
The U-Series logger requires an Onset-supplied USB interface cable to
connect to the computer. If possible, avoid connecting at temperatures
below 0°C (32°F)
or above 50°C (122°F).
1. Plug the large end of the USB interface cable into a USB port on the
computer.
2. Plug the small end of the USB interface cable into the side of the
logger as shown in the following diagram.
3. Load and use logger software to operate the logger (see software
manual).
Not Specified
+/- 5%
+/-3.5%
100%
90%
80%
70%
60%
50%
40%
30%
20%
U10 RH Accuracy
0
10
20
30
40
50
60
50
°C