Connecting the Sensor to a Station or Logger 

To connect the sensor to a station or logger, stop the station or 
logger from logging and insert the smart sensor’s modular jack 
into an available smart sensor port. See the station manual for 
details on operating stations or loggers with smart sensors. 

Mounting 

Use a solar radiation shield to mount the sensor. Use the cable 
clamps included with the RS3-B radiation shield to secure the 
smart sensor as shown. 

 

Temp/RH Sensor mounted in  

Solar Radiation Shield RS3-B (cross-section) 

Mounting Considerations 

 

A solar radiation shield is strongly recommended when 
measuring air temperature in direct sunlight. Solar 
radiation can be a significant source of error in the 
temperature and RH readings. 

 

When deploying an S-THC sensor, it is recommended that 
you mount the sensor vertically. If it must be mounted 
horizontally, then make sure the vent on the side of the 
sensor is vertical or facing down. 

 

Thermally isolate the sensor from the mounting surface to 
ensure accurate air temperature and humidity readings. 

 

Protect the sensor from direct exposure to the weather. 
This will prolong the sensor accuracy. 

 

Secure the smart sensor adapter to the mast with the 
cable ties as shown. Multiple smart sensor adapters can 
be stacked as shown in the example below on the right.  

 

 

 

Alternatively, mount the smart sensor adapter to a flat 
surface using two screws (no larger than a #6) and two 
washers as shown. 

 

 

If you are running sensor cables along the ground, use 
conduit to protect against animals, lawn mowers, 
exposure to chemicals, etc. 

 

Refer to the station manual and Tripod Setup Guide at 
www.onsetcomp.com/support/manuals for more 
information regarding setting up stations. 

Maintenance 

The temperature/RH smart sensor is sensitive to dust, salts and 
other airborne contamination. Periodically inspect the sensor 
probe and rinse it with distilled water to clean it. Do not use hot 
water, organic solvents, or detergents. Dry before use. 