Connecting the Sensor to a Station 

To connect the sensor to a station, stop the station from 

logging and insert the smart sensor’s modular jack into an 

available smart sensor port on the station. See the station 

manual for details on operating stations with smart sensors. 

Operation 

The PAR smart sensor supports measurement averaging. When 

measurement averaging is enabled, data is sampled more 

frequently than it is logged. The multiple samples are then 

averaged together and the average value is stored as the data 

for the interval. For example, if the logging interval is set at 10 

minutes and the sampling interval is set at 1 minute, each 

recorded data point will be the average of 10 measurements.  
Measurement averaging is useful for reducing noise in the data. 

It is recommended that you use measurement averaging 

whenever the PAR smart sensor is used in an area where the 

light level can vary quickly with respect to the logging interval 

(for example, under a partial plant canopy or during partly 

cloudy conditions). Note that fast sampling intervals less than 1 

minute may significantly reduce battery life. See the station 

manual for more details about battery life. 

Maintenance 

Dust on the sensor will degrade sensor accuracy. Periodically 

inspect the sensor, and if necessary, gently clean the diffuser 

with a damp sponge. Do not open the PAR smart sensor as 

there are no user serviceable parts inside. 

 Warning

:

 DO NOT use alcohol, organic solvents, abrasives, 

or strong detergents to clean the diffuser element on the light 

sensor. The acrylic material used in the light sensors can be 

crazed by exposure to alcohol or organic solvents. Clean the 

sensor only with water and/or a mild detergent such as 

dishwashing soap if necessary. It is recommended that you use 

vinegar to remove hard water deposits from the diffuser 

element. Under no circumstances should the smart sensor be 

immersed in any liquid. 

Verifying Sensor Accuracy 

It is recommended that you check the accuracy of the PAR 

smart sensor annually. The PAR smart sensor cannot be user-

calibrated. Onset uses precision components to obtain accurate 

measurements. If the sensor readings are slightly different than 

expected and you are using the sensor with an RX station, then 

you can use scaling in HOBOlink to perform a linear adjustment 

of the light sensor data to improve the accuracy. Use a light 

meter to take readings that you can use to determine a 

calibration factor to enter in HOBOlink. To do this: 
1.

 

Make sure the light meter is positioned at the same angle 

to the sun (or other light source) as the sensor (e.g. point 

the light meter straight up if the sensor is also pointing 

straight up). 

2.

 

Temporarily set the logging interval for the sensor in 

HOBOlink to 1 minute and Save. 

3.

 

Press the Connect button on the station. 

4.

 

Press the Start button if the station is not logging. 

5.

 

Make sure the station logs for at least 5 minutes so that 

there is a series of at least 5 readings (1 every minute). Note 

the light meter readings over the same time. 

6.

 

Divide the average light meter reading by the average PAR 

sensor reading to determine a calibration coefficient. 

7.

 

In HOBOlink, go to the device page and click   next to the 

sensor and enable scaling. 

8.

 

Enter the calibration coefficient from step 6 in the 

Multiplier field and leave the Offset field blank. 

9.

 

Fill in the Scaled Units with your preferred units, such as 

µ

or 

µ

mol/m

2

/sec. Type PAR in the Scaled Measurement Type 

field. 

10.

 

Save the changes. 

11.

 

Change the logging interval back to your preferred rate. 

If the smart sensor is not providing accurate data even after the 

adjustment for scaling, then it may be damaged or too far out 

of calibration. 