background image

HOBO MX Water Level Logger (MX2001-0x) Manual

 

U.S. and International Sales: 1-508-759-9500 9 

www.onsetcomp.com 

temperature rises above 85°F, the logger switches to the faster 
logging rate and records data every 30 seconds until the 
temperature falls back to 85°F. At that time, logging then 
resumes every 5 minutes at the normal logging interval. 
Similarly, if the temperature falls below 32°F, the logger 
switches to burst logging mode again and records data every 30 
seconds. Once the water temperature rises back to 32°F, the 
logger returns to Fixed mode, logging every 5 minutes. 

Note:

 

Sensor alarms, statistics, and the Stop Logging option 

Never 

Stop (Overwrites Old Data)

 are not available in burst logging 

mode. 

To set up burst logging: 

1.

 

Tap Devices. Tap the logger tile in the app to connect to the 
logger and tap Configure & Start. 

2.

 

Tap Logging Mode and then tap Burst Logging. 

3.

 

Select Low and/or High and enter  the low and/or high 
values. 

4.

 

Repeat step 4 for the other sensor if desired. 

5.

 

Set the burst logging interval, which must be faster than the 
logging interval. Keep in mind that the more frequent the 
burst logging rate, the greater the effect on battery life and 
the shorter the logging duration. 

Notes: 

 

Burst logging cannot be triggered by water level. It can be 
triggered only by temperature or differential pressure. 

 

Once you configure the logger, the logger checks the high 
and low burst limits once every 15 seconds. If you set the 
logging interval to less than 15 seconds and the sensor 
reading falls outside the levels, burst logging does not 
begin until the next 15-second refresh cycle. 

 

If you have configured high and/or low for more than one 
sensor, burst logging will begin when any high or low 
condition goes out of range. Burst logging does not end 
until all conditions on all sensors are back within normal 
range. 

 

The actual values for the burst logging limits are set to 
the closest value supported by the logger. 

 

Burst logging mode can begin or end when the sensor 
reading is within the resolution specifications. This means 
the value that triggers burst logging may differ slightly 
from the value entered. 

 

Once the high or low condition clears, the logging 
interval time is calculated using the last recorded data 
point in burst logging mode, not the last data point 
recorded in normal mode. For example, let’s assume the 
logger has a 10-minute logging interval and logged a data 
point at 9:05. Then, the high limit is surpassed and burst 
logging begins at 9:06. Burst logging then continues until 
9:12 when the sensor reading falls back below the high 
limit. Now back in normal mode, the next logging interval 
starts 10 minutes from the last burst logging point, or 
9:22 in this case. If burst logging had not occurred, the 
next data point would have been at 9:15. 

 

A New Interval event is created each time the logger 
enters or exits burst logging mode. See 

Logger Events

 for 

details on plotting and viewing the event. 

Statistics Logging 

During fixed logging, the logger records data for enabled 
sensors and/or selected statistics at the selected logging 
interval. Statistics are calculated at a sampling rate you specify 
with the results for the sampling period recorded at each 
logging interval. You can log the following statistics for each 
sensor: 

 

The maximum, or highest, sampled value 

 

The minimum, or lowest, sampled value 

 

An average of all sampled values 

 

The standard deviation from the average for all sampled 
values 

For example, a logger is configured with the absolute pressure, 
differential pressure, water temperature sensors enabled, and 
the logging interval set to 5 minutes. The logging mode is set to 
Fixed logging and all four statistics enabled and with a statistics 
sampling interval of 30 seconds. Once logging begins, the logger 
measures and records the actual absolute pressure, differential 
pressure, and water temperature sensor values every 5 
minutes. In addition, it calculates the water level and 
barometric pressure. The logger takes readings for the enabled 
sensors every 30 seconds and temporarily stores them in 
memory. The logger then calculates the maximum, minimum, 
average, and standard deviation using the samples gathered 
over the previous 5-minute period and log the resulting values. 
When downloading data from the logger, this results in 17 data 
series: 5 sensor-related series (differential pressure, absolute 
pressure, barometric pressure, temperature, and water level 
with data logged every 5 minutes) plus 12 maximum, minimum, 
average, and standard deviation series (four for differential 
pressure, four for absolute pressure, and four for temperature 
with values calculated and logged every 5 minutes based on the 
30-second sampling). Statistics are not available for barometric 
pressure and water level data. 

To log statistics: 

1.

 

Tap Devices. Tap the logger tile in the app to connect to the 
logger. Tap Configure & Start. 

2.

 

Tap Logging Mode and then select Fixed Logging Mode. 

3.

 

Tap the Statistics toggle to turn on Statistics. 

Important:

 You must select Fixed Logging Mode if you want to 

log water level and barometric pressure data. 

4.

 

Select the statistics you want the logger to record at each 
logging interval: Maximum, Minimum, Average, and 
Standard Deviation (average is automatically enabled when 
selecting Standard Deviation). Statistics are logged for all 
enabled sensors. The more statistics you record, the shorter 
the logger duration and the more memory is required. 

5.

 

Tap Statistics Sampling Interval and select the rate to use 
for calculating statistics. The rate must be less than, and a 
factor of, the logging interval. For example, if the logging 
interval is 1 minute and you select 5 seconds for the 
sampling rate, then the logger takes 12 sample readings 
between each logging interval (one sample every 5 seconds 
for a minute) and use the 12 samples to record the resulting 
statistics at each 1-minute logging interval. Note that the 
more frequent the sampling rate, the greater the effect on 
battery life. 

Summary of Contents for MX2001-0 Series

Page 1: ... 0 02 kPa 0 003 psi 0 21 cm 0 007 ft water Pressure Response Time 90 1 second at a stable temperature Pressure Absolute and Water Level Measurements MX2001 02 SS S Operation Range 0 to 400 kPa 0 to 58 psia approximately 0 to 30 6 m 0 to 100 ft of water depth at sea level or 0 to 33 6 m 0 to 111 ft of water at 3 000 m 10 000 ft of altitude Factory Calibrated Range 69 to 400 kPa 10 to 58 psia 0 to 4...

Page 2: ... 1 C at 25 C 0 18 F at 77 F see Plot A Response Time 90 5 minutes in water typical Stability Drift 0 1 C 0 18 F per year Logger Operating Range 20 to 50 C 4 to 122 F Radio Power 1 mW 0 dBm Transmission Range Approximately 30 5 m 100 ft line of sight Wireless Data Standard Bluetooth Low Energy Bluetooth Smart Logging Rate 1 second to 18 hours Logging Modes Fixed interval multiple intervals with up ...

Page 3: ... to achieve full temperature compensation of the pressure sensor There can be up to 0 5 of additional error due to rapid temperature changes Measurement accuracy also depends on temperature response time Note Read all the directions in the Deploying a Logger section before deploying the logger Logger Components and Operation Disassembled top end The MX water level logger consists of a top end and ...

Page 4: ...layed as when the logger is stopped 6 Deploy the logger if ready If you are ready to deploy the logger place it in the deployment location see Deploying the Logger for deployment guidelines Otherwise skip deployment and complete the following steps as a test 7 Connect to the logger Tap the logger tile in the app to connect to the logger Tips for connecting Make sure the top end of the logger is wi...

Page 5: ...he current water level when the logger starts logging make sure the sensor is in position when logging starts Make this reading as accurate as possible since this reading is used to calibrate the water level data and adjust it to your reference See section 5 starting on page 7 for information on how to measure and enter the reference water level The logger must be logging for you to enter the refe...

Page 6: ...grommets are available at www onsetcomp com part number GROMMET 5 To install the grommet with the logger 1 Drill two 3 mm or 1 8 inch holes on opposite sides of the well cap for cross ventilation The holes should be just above the top of the grommet when installed as shown 2 Install the grommet with the upper ring sitting on top of the well as shown 3 Install the well cap on top of the well refer ...

Page 7: ...the top end unit in direct sun as that can lead to more rapid temperature swings which can lead to more measurement error and possibly hurt logger reliability To deploy the logger 1 Make sure the top end is connected to the sensor end with the water level logger cable as described in Getting Started with the Logger 2 Attach any small parts such as bolts with nuts before you go into the field so yo...

Page 8: ... alarm is set to 85 F and the duration is set to 30 minutes then the alarm will trip once the sensor readings have been above 85 F for a total of 30 minutes since the logger was configured Consecutive The alarm trips once the sensor reading is out of the acceptable range continuously for the selected duration For example the high alarm is set to 85 F and the duration is set to 30 minutes then the ...

Page 9: ... Logger Events for details on plotting and viewing the event Statistics Logging During fixed logging the logger records data for enabled sensors and or selected statistics at the selected logging interval Statistics are calculated at a sampling rate you specify with the results for the sampling period recorded at each logging interval You can log the following statistics for each sensor The maximu...

Page 10: ...wnloads data to the phone tablet or computer 3 Once the download is complete tap Done to return to the previous page or tap Export and Share to save the file in the specified format 4 When the export file has successfully been created tap Done to return to the previous page or tap Share to use your device s usual modes of sharing You can also upload data automatically to HOBOlink Onset s cloud sof...

Page 11: ...All pressure sensors drift over time The drift for the pressure sensors and electronics in the logger is less than 0 5 FS worst case per year In most applications drift is not a significant source of error because the offset created by any drift is zeroed out when you enter the reference water level in the app at the beginning of each deployment In effect you are re zeroing the sensor each time yo...

Page 12: ...rcuit board is normal when the batteries are installed Reinsert the board 6 Inspect the gore vent on the end cap to make sure it is not damaged Contact Onset or an Onset distributor if you need a replacement vent 7 Inspect the O rings and the mating surface on the logger for any damage clean off any debris and re lubricate the O rings with the grease You can leave the O rings on the end cap during...

Page 13: ...cations not expressly approved by the party responsible for compliance could void the user s authority to operate this equipment Industry Canada Statements This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may...

Reviews: