Hobo International PENDANT G UA-004-64 User Manual Download Page 3

HOBO Pendant G Acceleration Data Logger 

3

 

Operation 

A light (LED) on the front of the logger confirms logger operation. 
The following table explains when the light blinks during logger 
operation. 

When: The 

light: 

The logger is awaiting an 
Interval, Delayed, or Trigger 
Start  

Blinks once every eight seconds until 
logging begins. 

Logger is subjected to a 
magnetic field 

Blinks once every second when the 
magnet is present, then (Normal mode 
only) blinks rapidly three times when 
logging begins. 

The logger is logging in 
Normal mode 

Blinks once every one to four seconds 
(the shorter the logging interval, the 
faster the light blinks); blinks when 
logging a sample 

The logger is logging in Fast 
mode 

Blinks once every quarter-second 

Mounting the logger 

The provided mounting bracket offers two orientation options for 
securing the logger to a rigid surface.  

Secure the bracket to the object of interest, then use the mounting 
screw provided to attach the logger to the bracket. 

    

  

 

Protecting the logger 

The logger can be damaged if the water depth rating is exceeded. The 
depth rating is approximately 30 m (100 ft) at temperatures below 
20°C (68°F), but is less in warmer water. Refer to Plot B for details. 

Do not store the logger in the coupler. Remove the logger from the 
coupler when you are not using it. When the logger is in the coupler 
or near a magnet, it consumes more power and will drain the battery 
prematurely. 

Keep the logger away from magnets. Being near a magnet can cause 
false coupler events to be logged. It can also launch the logger 
prematurely if it is waiting for a trigger start. 

Note! Static electricity may cause the logger to stop logging. 

To 

avoid electrostatic discharge, transport the logger in an anti-static 
bag, and ground yourself by touching an unpainted metal surface 
before handling the logger. For more information about electrostatic 
discharge, visit http://www.onsetcomp.com/support/support.html. 

Periodically inspect the desiccant and dry it if it is not bright blue. 
The desiccant pack is located in the cap of the logger. To dry the 
desiccant, remove the desiccant pack from the cap and leave the pack 
in a warm, dry location until the bright blue color is restored. (Refer 
to the “Battery” section for instructions on removing and replacing 
the logger cap.) 

Temperature range 

Desiccant maintenance 
schedule 

Less than 30°C (86°F) 

Approximately once per year  

30° to 40°C (86° to 104°F)  Approximately every six 

months 

Over 40°C (104°F) 

Approximately every three 
months 

 

Factory calibration 

The logger is factory-calibrated and does not require any additional 
calibration. To verify proper calibration, check the logger’s status 
(when it is not logging) in the HOBOware software to check the 
logger’s response to gravity. When any of the axes are held 
horizontally, you should see a reading close to zero. When held 
vertically, with the arrow facing downwards, the response should be 
close to 1 

g

.  

Battery 

The logger requires one 3-Volt CR-2032 lithium battery. Battery life 
varies based on the temperature and the frequency at which the 
logger is recording data (the logging interval). A new battery 
typically lasts one year with logging intervals greater than one 
minute. Deployments in extremely cold or hot temperatures, or 
logging intervals faster than one minute, may significantly reduce 
battery life. Continuous logging at the fastest logging rate of 0.01 
seconds (100 Hz) will deplete the battery in as little as one week. 

To replace the battery: 

1.  Remove the two screws that secure the end cap to the case and 

remove the cap. 

2.  Examine the desiccant pack that is tucked into the cap. If the 

desiccant is not bright blue, put the desiccant pack in a warm, dry 
place until the blue color is restored. 

3.  Gently tap the case to loosen the circuit board and remove it from 

the case.  

 

4.  Carefully push the battery out of the holder with a small, 

nonmetallic blunt instrument. 

5.  Insert a new battery, positive side facing up. 

6.  Return the circuit board and label to the case, carefully aligning 

the circuit board with the grooves in the case so that the battery 
faces the ridged side of the case.  

Reviews: