background image

Copyright HK Instruments 2019

www.hkinstruments.fi

Installation version 8.0 2019

The seller is obligated to provide a warranty of five years for the deliv-
ered goods regarding material and manufacturing. The warranty period 
is considered to start on the delivery date of the product. If a defect in 
raw materials or a production flaw is found, the seller is obligated, when 
the product is sent to the seller without delay or before expiration of 
the warranty, to amend the mistake at his/her discretion either by re-
pairing the defective product or by delivering free of charge to the buy-
er a new flawless product and sending it to the buyer. Delivery costs 
for the repair under warranty will be paid by the buyer and the return 
costs by the seller. The warranty does not comprise damages caused by 
accident, lightning, flood or other natural phenomenon, normal wear 
and tear, improper or careless handling, abnormal use, overloading, im-
proper storage, incorrect care or reconstruction, or changes and instal-
lation work not done by the seller or his/her authorized representative. 
The selection of materials for devices prone to corrosion is the buyer’s 
responsibility, unless otherwise is legally agreed upon. Should the man-
ufacturer alter the structure of the device, the seller is not obligated to 
make comparable changes to devices already purchased. Appealing for 
warranty requires that the buyer has correctly fulfilled his/her duties 
arisen from the delivery and stated in the contract. The seller will give 
a new warranty for goods that have been replaced or repaired within 
the warranty, however only to the expiration of the original product’s 
warranty time. The warranty includes the repair of a defective part or 
device, or if needed, a new part or device, but not installation or ex-
change costs. Under no circumstance is the seller liable for damages 
compensation for indirect damage.

SELECTING THE MEASUREMENT RANGE 

CONTINUED

Chart 2 - Inlet A

Range 7

Range 8

Range 1

Range 2

Range 3

Range 4

Range 5

Range 6

Jumper J1
Jumper J2
Jumper J3

(Grey colour indicates that a jumper is installed. Reference Figur

e 3 and Schematics for jumper installation.)

3) For inlet B: Install jumpers on J4, J5 and J6 as required.
a. Using the range number from 1c above, find the corresponding 

range number in Chart 3.

b. Install jumpers on J4, J5 and J6 on device, as shown under the range 

number in Chart 3. (Grey colour indicates that a jumper is installed. 
Reference figure 3 for jumper installation.)

Chart 3 - Inlet B

Range 7

Range 8

Range 1

Range 2

Range 3

Range 4

Range 5

Range 6

Jumper J4
Jumper J5
Jumper J6

(Grey colour indicates that a jumper is installed. Reference Figur

e 3 and Schematics for jumper installation.)

STEP 6: SELECTING THE RESPONSE TIME

The response time affects how fast the transmitter reacts to changes in 
the system. The response time is the time the device takes to reach 63 
% of the measured value. To smooth out unstable pressure fluctuations 
in airflow applications, select a longer response time.

Example:
Selected response time: 4.0 seconds
Result: Output signal achieves a new value in 20 seconds (Response 
time*5)

To change response time, install or remove jumper on J7.
(see Figure 3)
1) Install jumper on J7 for 4.0 second response time.
2) Remove jumper from J7 for 0.8 second response time.

NOTE: This will change the response time for both differential pressure 

measurements.

WARRANTY POLICY

STEP 7: ZEROING THE DEVICE

NOTE! Always zero the device before use.

Manual Pushbutton zero point calibration:
NOTE: Supply voltage must be connected at least one hour prior to 
zero point adjustment.
a) Disconnect both pressure tubes from the pressure ports l 

and –.

b) Push down the zero button until the LED light (red) turns on and 

the display reads “zeroing” (display option only). (see Figure 4)

c) The zeroing of the device will proceed automatically. Zeroing is 

complete when the LED turns off, and the display reads 0 (display 
option only).

ZEROING THE DEVICE

 CONTINUED

d) Reinstall the pressure tubes ensuring that the High pressure tube 

is connected to the port l, and the Low pressure tube is 
connected to the port labeled −.

Figure 4

a) Disconnect both pressure tubes from the pressure ports l 
and –.

b) Push down the zero button (joystick) until the LED light (red) 
turns on and the display reads “zeroing” (display option only). (see 
Figure 6)
c) The zeroing of the device will proceed automatically.  Zeroing is 
complete when the LED turns off, and the display reads 0 (display 
option only).  
d) Reinstall the pressure tubes, ensuring the High pressure tube is 
connected to the port l, and the Low pressure tube is 
connected to the port labeled −.

Press down button

LED turns ON

Display reads “Zeroing”

LED

RECYCLING/DISPOSAL

The  parts  left  over  from  installation  should  be 
recycled  according  to  your  local  instructions. 
Decommissioned  devices  should  be  taken  to  a 
recycling site that specializes in electronic waste.

Summary of Contents for DPT-DUAL Series

Page 1: ...from 0 7000 Pa DPT Dual series devices are commonly used in HVAC R systems for fan blower and filter monitoring pressure and flow monitoring valve and damper control SPECIFICATIONS Performance Accuracy from applied pressure Model 2500 Pressure 125 Pa 1 2 Pa Pressure 125 Pa 1 1 Pa Model 7000 Pressure 125 Pa 1 5 2 Pa Pressure 125 Pa 1 5 1 Pa including general accuracy linearity hysteresis long term ...

Page 2: ... device is now ready for configuration WARNING Apply power only after the device is properly wired 1 Select the mounting location duct wall panel 2 Use the device as a template and mark the screw holes 3 Mount with appropriate screws Figure 1b Mounting orientation Figure 1c Application connections Display connection Output 0 10 V Output 0 10 V 24 V GND LED Zero button Terminal block Pressure senso...

Page 3: ...talled Circuit closed Grey colour indicates that a jumper is installed Chart 1 Model DPT Dual 2500 Range 1 Range 2 Range 3 Range 4 Range 5 Range 6 Range 7 Range 8 Pa kPa mbar inchWC 100 100 0 10 0 10 1 00 1 00 0 40 0 40 0 100 0 0 10 0 1 00 0 0 40 0 250 0 0 25 0 2 50 0 1 00 0 500 0 0 50 0 5 00 0 2 00 0 1000 0 1 00 0 10 0 0 4 00 0 1500 0 1 50 0 15 0 0 6 00 0 2000 0 2 00 0 20 0 0 8 00 0 2500 0 2 50 0...

Page 4: ... shown under the range number in Chart 3 Grey colour indicates that a jumper is installed Reference figure 3 for jumper installation Chart 3 Inlet B Range 7 Range 8 Range 1 Range 2 Range 3 Range 4 Range 5 Range 6 Jumper J4 Jumper J5 Jumper J6 Grey colour indicates that a jumper is installed Reference Figure 3 and Schematics for jumper installation STEP 6 SELECTING THE RESPONSE TIME The response ti...

Reviews: