DENSITY TEST 

Step 1: Press Setting sample density and enter the density value. 

 

 

Step 2: Press weighing in the air,then put the sample on the top of density kit. 

   

When the weighing digit value stable, press again then will be show green square. 
Means confirm the weight in the air. 
 

10 

Summary of Contents for BTH Series

Page 1: ... 5 ...

Page 2: ... 5 I9 57 F5A ...

Page 3: ...RES 01 1 2 APPLICATIONS 01 2 THE WEIGHING MODE 02 2 1 KNOW YOUR BALANCE 02 2 2 WEIGHING MODE 02 2 3 SYSTEM SETTING 12 2 4 CALIBRATION 13 2 5 WEIGHING SYSTEM 15 3 PRINT 17 3 1 DATA OUTPUT 17 3 2 PRINTER SETTING 17 4 TROUBLE SHOOTING 20 5 DISPLAY CALIBRATION 21 ...

Page 4: ...2 Features Stainless Steel Pan Super 7 inch HD Touch panel display RS232 RJ45 USB Interface Mains adapter supplied as standard Height adjustable feet Internal Auto Calibration Optional Selectable measure units mg g oz ct Memory for accumulated time 1 3 Applications Weighing Net weight tare Under weighing Piece counting function Density Test Percentage Test Check Weighing Accumulative total Output ...

Page 5: ...s Please read this Manual thoroughly before starting the operations If you need any clarifications feel free to contact us 2 2 Weighing Mode The BHT scale are ideal for laboratory and general purpose weighing The scale can also be used for some advanced weighing functions WEIGHING Standard weighing mode CHECKING Set the upper and lower limit for checkweigh function PECENTAGE Compare the sample wit...

Page 6: ...pacity and resolution 5 Weighing value 6 Note color blue mean normal red means overload 7 Key pad 8 Auxiliary tools Calibration Mode choose Language The weighing function just use for weighing system The sample weight will direct Showing in this page And we can export the result in output and input page 03 ...

Page 7: ...nd resolution 5 Weighing value 6 Note color blue mean normal red means overload 7 Key pad 8 Auxiliary tools Calibration Mode choose Language 9 Tare weight Put the container press tare the tare weight will show here 10 G W Gross weight 11 Total weight and test number 12 Upper and Lower limit setting 04 ...

Page 8: ...Upper and Lower Setting Step 1 press the Upper or Lower digit Step 2 Press The Enter Input box Step 3 Enter the digit for upper limit The lower limit use same steps 05 ...

Page 9: ...eans overload 8 Auxiliary tools Calibration Mode choose Language 9 Key pad 10 Confirm mark Step 1 Press 100 then put on the weight when it goes stable press again It will mark a green square so means 100 set success Step 2 Press XX then put on the weight when it goes stable press again It will mark a green square so means xx confirm Step 3 Press it will come out the result 06 ...

Page 10: ...tatus Bar Battery Stable Speed 2 Weighing mode choose 3 Date and time 4 Capacity and resolution 5 Weighing value 6 Quantity 7 Sample weight 1 pcs 8 Auxiliary tools Calibration Mode choose Language 9 Counting details 10 Key pad 07 ...

Page 11: ...tep 1 Put on the sample weight Step 2 Press the quantity And enter the sample quantity Steps 3 Back to the counting page and start the counting test We also can direct enter the sample weight and use counting functions 08 ...

Page 12: ...id weight in the liquid 6 Setting the liquid density for solid test 7 Note color bar blue mean normal red means overload 8 Auxiliary tools Calibration Mode choose Language 9 Key pad 10 Confirm mark 11 The density test result This density function need use our density kits or under weighing function for density test Density hook 09 ...

Page 13: ...mple density and enter the density value Step 2 Press weighing in the air then put the sample on the top of density kit When the weighing digit value stable press again then will be show green square Means confirm the weight in the air 10 ...

Page 14: ...id then put the sample inside the liquid when the weighing digit stable press again then will be show green square Means confirm the weight in the liquid Step 4 Press the density the result will be come out the density unit is g cm3 11 ...

Page 15: ... We can set the system parameters in system setting function The setting items include as blew 1 weighing speed one means Fast 3 means slow 2 stability one means high 3 means low 3 language 4 Printer 5 Date and time setting 12 ...

Page 16: ...ation 1 Press one point calibration please operate follow the note 2 It will display the value for one point calibration and show put on the weight then put on the same value weight 3 When it displays move away the weight then remove the weight 4 Waiting it go zero Calibration complete 5 If choose internal calibration please use the automatic calibration in the weighing page 13 ...

Page 17: ...IBRATION key then operate as the note which will show in the touch screen there will be show 3 point calibration value for this multi calibration 2 After calibration you can check the liner with your standard weight 2 4 3 Internal Setting This function is setting the internal calibration period operator can choose the automatic calibration time for itself calibration We can set the time what we ne...

Page 18: ... for make sure whether the hardware suitable or not 2 5 1 CAPACITY AND RESOLUTION SETTING We can change the capacity and resolution for each scale Need Password 1 Press CAPACITY RESOLUTION key enter the setting 2 Press enter blank then enter the capacity value you want Press confirm 3 Press BACK complete setting 15 ...

Page 19: ...ibration weight We advise you choose calibration point 20 Capacity it can be reduced error you d better choose the full range as one point calibration 2 MAX Setting the full capacity weight Keep first number 6200g will set 6000g 3 HALF Setting 50 weight value Example capacity 6200g we set 50 3000g 4 MIN Setting small point 30 capacity the better way is one weight can meet require for example 6200g...

Page 20: ...se the output or input you want Step 4 There will be note you output or input complete 2 5 4 AUXILIARY KEY SETTING We can set the CAL key function for 1 External calibration or 2 Internal calibration 3 PRINT 3 1 Data Output 1 Model or a decimal point 2 A space or a decimal point 3 A space or 4 or or a decimal point 5 data 6 Data or a decimal point 7 Data or a decimal point 8 Data or a decimal poin...

Page 21: ...or different print or connect with the computer There are five kinds of print format for label print we can choose the mode then enter the information which the operation need And we also can choose the baud rate and communication which we need Step 1 Choose the mode which you use 18 ...

Page 22: ...Step 2 Choose the suitable baud rate and communication mode Step 3 Enter the operator information Press Confirm can enter the input information 19 ...

Page 23: ...ble Bad working conditions The wind screen is open Something between the table and balance The power unstable Weighing unstable Keep the environment stable close the windows and doors Close the glass door Take away the things Connect the stable power The weighing digits is wrong The balance not calibration Not tare before weighing No adjust the level Calibration Tare before weighing Adjust the lev...