15
v02.1110
Order On-Line: www.hittite.com. For technical application questions: Phone: 978-250-3343 or [email protected]
HMC-T2200 Synthesized Signal Generator Family User Manual
10000000;40000000000
pow? min; pow? max
-60.0;30.0
swe:dwel? min; dwel? max
0.000300;4294.967044
freq:resolution?
1
diag:comp T2100
<< power cycle >>
*idn?
Hittite,HMC-T2100,000010,2.2 4.4
freq? min; freq? max
10.00e6;20000.00e6
pow? min; pow? max
-36.0;27.0
swe:dwel? min; dwel? max
0.000100;4294.967044
freq:resolution?
0.01e6
diag:compatibility?
T2100
3.3.5
DIAGnostic:MACaddress?
Read the MAC Address from the unit.
MAC addresses are used for Ethernet communication. The MAC Address is programmed at the fac-
tory, and cannot be changed, but it may be helpful to read back for debug purposes.
Your router may also have a utility which will tell you the MAC address of everything connected to it.
See Also:
SYSTem:COMMunicate:ETHernet:ADDRess,SYSTem:COMMunicate:ETHernet:DHCP
3.3.5.1
Example: Read theMAC Address from a HMC-T2240
DIAG:MAC?
00242a987654
3.3.6
DISPlay:ENABle[?] - LCD ON/OFF
DISPlay:ENABle OFF causes the following to appear on the front panel LCD:
DISPLAY IS
CURRENTLY OFF
The display can be turned back on with
DISP: ENAB ON.
The query form is
DISPlay:ENABle? which returns 1 or 0.
The reset (
*RST) state is ON. Because the SECurity commands are not implemented, the dis-
play will go to a different state whenever
*RST is issued, per SCPI-99. However the reset state