background image

130-0047 Issue 8

USED FOR CONTROL, OR THE ASSOCIATED CONTROL LOOP
SHOULD BE DISABLED BEFORE COMMENCING.

5.1 Sensor life

The N sensors have a life of 100,000 oxygen % hours or 18 months
- whichever is the sooner; and the L type a life of 16,000 oxygen %
hours or 18 months whichever is the sooner.  Each sensor is date
labelled (see the Storage of the Measuring Sensor section 6.2 for
the code) when supplied. This date should be used to establish the
expiry date.

5.1.1 Sensor failure modes

All known sensor failure modes result in a loss or lowering of output.
Thus applications that look for oxygen depletion are automatically
fail safe and vice-versa

.

5.2 Calibration interval

It is recommended that the calibration is verified every month.  This
frequency of verification is for typical industrial applications.  When
the application is critical, the frequency of verification should be
increased in line with local safety standards.

5.3 Calibration gas level

The level of oxygen in the calibration gas depends on the particular
model and in particular the type of sensor fitted.  In general, for
optimum accuracy, it best to calibrate with a gas with the same
composition and oxygen concentration as the normal sample. If you
calibrate on air, the response time for the sensor to stabilise back to
its normal working level after being exposed to air should be taken
into account.
Moreover if an air calibration is performed, it should be borne in
mind that most air taken from the atmosphere is moist.  The normal
figure of 20.95% for the oxygen content of air is for dry air.  Ambient
air contains nearer 20.7% because of dilution effect of the

Page 12 of 18

moisture.  It is this figure that should be used when calibrating in this
way.

5.4 Calibration of a percent range instrument.
      

(See section 5.5 for instruments scaled in PPM)

5.4.1 Piping

The piping carrying the calibration gas must not have any leaks, and
should have an outlet pipe at least 300mm long to reduce any effect
of back diffusion.  The flow rate of  the gas must be controlled and
should be the same as the flow rate of the sample from the process.
Also, the pressure at the sample outlet must be the same as that
when the analyser is measuring the sample.  This avoids any errors
due to the  pressure at the sensor.  Refer to Fig 2 for details of a
typical installation.

5.4.2 Calibration method

Minutes

Response of an N type sensor from an air saturated 

condition when zero grade nitrogen is applied.

Fig. 4

10

2

3

Hours

PP

M

 O

x

yge

n

4

5

6

7

8

9

10

11

12

20

30

40

50

60

70

80

90

100

110

120

130

140

150

160

170

Fig. 5

Response of an L type sensor from an air saturated 

condition when zero grade nitrogen is applied

Summary of Contents for G1010 Series

Page 1: ...nded by 93 68 EEC by the application of the following standards EN50081 Part1 1993 Generic Emission Standard EN50082 Part1 1995 Generic Immunity Standard Installation Requirements for EMC To ensure co...

Page 2: ...number and we will be pleased to assist Page 1 of 18 2 0 SPECIFICATION Display Multi digit LCD character height 12 7mm Display ranges 1 version N type sensor Display range 0 to 55 Resolution 0 1 from...

Page 3: ...1 and 5 This instrument has been designed to meet the requirements of the EMC Directive 89 336 EEC and the requirements of the Low Voltage Directive 73 23 EEC when installed in accordance with these i...

Page 4: ...easy connection of a standard gas for calibration checks It is particularly important that good pipework connections are made when low levels of oxygen are being measured Page 5 of 18 For all parts pe...

Page 5: ...on The bung should be removed from the sensor and the sensor coupled to the holder as rapidly as possible THIS ASSUMES THAT THE SAMPLE THAT WILL INEVITABLY ESCAPE WILL NOT CONSTITUTE ANY SORT OF HAZAR...

Page 6: ...130 0047 Issue 8 Page 8 of 18...

Page 7: ...lowly and the least significant number can be changed by Page 9 of 18 giving the buttons a short click If the button is kept depressed the rate accelerates rapidly to enable large changes in value to...

Page 8: ...hanges of oxygen concentration within the instrument s span Short time excursions to high levels of oxygen such as may be experienced during calibration are recovered from in a matter of a few minutes...

Page 9: ...ir the response time for the sensor to stabilise back to its normal working level after being exposed to air should be taken into account Moreover if an air calibration is performed it should be borne...

Page 10: ...es that entrain air oxygen and can take several hours to purge down For this reason they should be avoided Calibration gas level for L sensors Ideally the calibration gas level should be as near to th...

Page 11: ...TC IF IT ESCAPES INTO THE ATMOSPHERE IT IS THE RESPONSIBILITY OF THE USER TO ENSURE THIS IF A HAZARD WILL RESULT THEN THE METHOD SHOULD BE MODIFIED ACCORDINGLY CONSULT HITECH OR THEIR REPRESENTATIVES...

Page 12: ...ech Instruments Ltd or their local representative for repair When ordering spare sensors or raising queries on the instrument it is important that the serial number or job number is quoted These numbe...

Reviews: