-27-
MN1-1162 Rev. 21
5-2. Reprocessing instruction according to ISO 17664
Take care about clean circumstances before using the probe on the next patients. If processors reprocess this
equipment, refer to these instructions.
Table 1
WARNINGS
・
The probe is delivered unsterile. Prior to the first use, reprocess the probe.
・
Do not exceed 60 °C [140 °F].
・
Probe connector has no water resistance.
Limitations on
reprocessing
The probe is not completely submergible (Do not immerse the probe into any liquid
beyond the range of IPX7. The range is indicated in the section 2-2 "specification".
)
Parts which are not submergible can only be disinfected by wipe disinfection.
Transportation
before using
Sterile pouch or container should be kept between transportation from Central
Sterile Supply Department (CSSD) to operating room. Be careful that no damages
are applied to sterile pouch or container for transportation.
The level of processing required depends on the type of equipment and its use.
The CDC (Centers for Disease Control and Prevention) in the USA and the RKI (Robert Koch Institute) in
Germany classify medical devices according to their use. For each classification, they specify the level of
disinfection/sterilization processing that is required before use. Table 3 summarizes this information.
Table 2
Classification
Definition
Processing
Noncritical
Application part only contacts
intact and uninjured skin
Cleaning
↓
Disinfection
[in the USA, low-level disinfection]
Semicritical
Application part contacts mucosa
(intracavitary application)
Cleaning
↓
Disinfection (Disinfectant with bactericidal,
fungicidal and virucidal effect)
[in the USA, high-level disinfection or
sterilization]
Critical
Application part contacts
intracorporeal tissue directly
(intraoperative application)
Cleaning
↓
Disinfection
↓
Sterilization *
1
*
1.
When sterilization is not possible, the FDA in the USA recognize that disinfection (in the USA, high-
level disinfection) and the use of a sterile gel and sterile transducer cover, as described in the instructions
provided with the transducer cover, is an accepted method of infection control for probe.
Summary of Contents for UST-9124
Page 2: ...MN1 1162 Rev 21 ii ...
Page 6: ...MN1 1162 Rev 21 vi ...
Page 14: ... 8 MN1 1162 Rev 21 Label A 2 Storage case Label B ...
Page 22: ... 16 MN1 1162 Rev 21 ...
Page 30: ... 24 MN1 1162 Rev 21 ...
Page 46: ... 40 MN1 1162 Rev 21 ...
Page 48: ... 42 MN1 1162 Rev 21 ...
Page 50: ... 44 MN1 1162 Rev 21 ...
Page 54: ... 48 MN1 1162 Rev 21 ...